The article focuses on the effect of transpedicular cemented and non-cemented screws applied in osteoporotic vertebrae and on the stability of connection between the bone and screw when subjected to extraction test. Cemented and non-cemented screws were inserted in experimental vertebral samples and extracted using a standard testing device. The obtained results prove that the extraction force in cemented screws is significantly higher than in non-cemented screws (t-test, p-value = 0.007).
Příspěvek se zabývá posouzením transpedikulárních cementovaných a necementovaných šroubů aplikovaných do kadaverózního osteoporotického obratle; pevnost spojení kost - šroub jsme hodnotili při vytrhávací zkoušce. Do zkušebních vzorků obratlů byly aplikovány cementované a necementované šrouby a na zkušebním stroji vytrženy. Naměřené výsledky potvrzují, že vytrhávací síla je pro cementované šrouby statisticky významně vyšší než pro necementované (t-test, p-value = 0,007).
Słowa kluczowe
Rocznik
Tom
Strony
75--78
Opis fizyczny
Bibliogr. 1 poz
Twórcy
autor
- Centre of Mechatronics - Institute of Thermomechanics Academy of Sciences of the Czech Republic and Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, Brno University of Technology
autor
- Spinal Unit, Trauma Hospital of Brno, Ponavka 6, 662 50 Brno, Czech Republic
autor
- Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, Brno University of Technology
autor
- Spinal Unit, Trauma Hospital of Brno, Ponávka 6, 662 50 Brno, Czech Republic
autor
- Centre of Mechatronics - Institute of Thermomechanics Academy of Sciences of the Czech Republic and Institute of Solid Mechanics, Mechatronics and Biomechanics, Faculty of Mechanical Engineering, Brno University of Technology
Bibliografia
- 1. Seedbeck J., Goldhahn J., Morlock M.M., Schneider E., Mechanical behavior of screws in normal and osteoporotic bone, Osteoporos Int (2005) 16: S107-S111.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-3a6a3b08-22a5-4224-bc79-57b82b363cbc