This paper presents the concept, design and experimental results of tests, in laboratory conditions, of a 2-port, space-saving pneumatic probe. The main features of the probe are a result of restrictions imposed by the functioning environment: a limited space for probe installation, the possibility of processing only two output signals and requirements for simplicity and reliability of the probe. A cylindrical shape of the probe tip is proposed as a general concept, similar to the classic, cylindrical, 3-port probe. The main difference arises in using only two pressure signals: one from the overpressure zone on the front side of the cylinder and another, from the underpressure zone on the back side. After performing an appropriate calibration procedure, it is possible to measure the flow velocity and correct the pressure difference obtained by means of a velocity coefficient kv. This paper also presents an analysis of the kv coefficient uncertainty to evaluate the quality of measurements.
2
Dostęp do pełnego tekstu na zewnętrznej witrynie WWW
W artykule przedstawiono wyniki wzorcowań ośmiu sond pneumatycznych typu S i dokonano oceny ich właściwości metrologicznych. Dla przebadanych sond stwierdzono stosunkowo duży rozrzut wartości współczynnika pozwalającego związać mierzoną różnicę ciśnień z wartością prędkości przepływu. Uzyskane różnice wskazują na brak możliwości przeniesienia wyników wzorcowań pomiędzy różnymi sondami tego typu - niezbędne jest wzorcowanie każdego z egzemplarzy osobno. Dokonano także analizy możliwych przyczyn uzyskanej rozbieżności parametrów poszczególnych sond.
EN
The paper presents the calibration results for eight different S-type pneumatic probes as well as the evaluation of their metrological features. A large spread of the coefficient - allowing to relate the measured pressure difference to the flow velocity - was found in evaluating the probes. The differences found indicate that calibration results are not transferrable between different probes of this type and that probes need to be calibrated individually. Potential reasons for the spread of the parameters obtained for different probes were also examined.
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.