An adsorptive stripping voltammetric procedure for the determination of trimethoprim at an in situ plated lead film electrode on glassy carbon substrate was described. The lead film plating process and accumulation of trimethoprim were performed simultaneously in an acetate buffer solution of pH = 5.8 and at potential -1.275 V. The measurements were carried out in undeaerated solutions. The calibration graph, for an accumulation time of 60 s, was linear in the range from 1 x 10-8 to 2x10-7 mol L-1. The detection limit was 3.5 x 10-9 mol L-1: the relative standard deviation for 1 x 10-7 mol L-1 trimethoprim was 4.2%. The proposed procedure was successfully applied to trimethoprim determination in pharmaceutical preparations and human urine samples.
PL
W pracy opisano procedurę oznaczania trimetoprimu na tworzonej in situ błonkowej elektrodzie ołowiowej na podłożu z węgla szklistego, metodą adsorpcyjnej woltamperometrii stripingowej. Tworzenie bionki ołowiu i zateżanie trimetoprimu przeprowadzano jednocześnie, z roztworu buforu octanowego o pH = 5,8, przy potencjale -1.275 V. Pomiary wykonywano z roztworów nieodtlenionych. Krzywa kalibracyjna przy czasie zatężania 60 s jest liniowawzakresieod l x 10-8 do 2x 10-7 mol L-1. Granica wykrywalności wynosi 3.5 x 10-9 mol L-1 a względne odchylenie standardowe, przy oznaczaniu trimetoprimu o stężeniu l x 10-7 mol L-1 wynosi 4,2%. Zaprezentowana procedura została wykorzystanado oznaczania trimetoprimu w preparatach farmaceutycznych i próbkach ludzkiego moczu.
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.