The Jyros valve has unique hydrodynamic characteristics comparing with other types of valves. In this study, precise hydrodynamic characteristics of the valve were examined using a mock circulatory system, which we have developed. Further, the influence of leaflet rotation is shown to reduce flow resistance under the rotational flow condition which was produced by a spiral vortex pump. However, hemolysis was not revealed to be reduced by the Jyros.