Ten serwis zostanie wyłączony 2025-02-11.
Nowa wersja platformy, zawierająca wyłącznie zasoby pełnotekstowe, jest już dostępna.
Przejdź na https://bibliotekanauki.pl

PL EN


Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników
2013 | Vol. 33, no. 3 | 156--163
Tytuł artykułu

The effect of type of filters and collimators on the SPECT axis of rotation error

Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
High resolution and artifact free SPECT reconstruction of the distribution of radioactive sources over the entire imaging field requires exact alignment between the electronic and the mechanical axis of rotations. In this paper, the effect of misalignment between the electronic and the mechanical axis of rotations of SPECT system equipped with low-energy high-resolution and low-energy general-purpose collimators through the field of view has been studied. In addition, the useful dimensions of the SPECT field of view without artifact due to axis of rotation error have been calculated. Furthermore, the roles of digital filters in reducing this effect have been studied.
Słowa kluczowe
Wydawca

Rocznik
Strony
156--163
Opis fizyczny
Bibliogr. 15 poz., wykr.
Twórcy
autor
  • Physics Department, University of Guilan, Namjoo Street, PO Box: 41365-1159, Rasht, Guilan, Iran, Hammed@guilan.ac.ir
Bibliografia
  • [1] Khalil M. Basic Sciences of Nuclear Medicine. Berlin: Springer; 2011.
  • [2] Saha GB. Fundamentals of Radiopharmacy, 5th Edition, Berlin: Springer; 2004.
  • [3] Sharp PF, Gemmell HG, Murray AD, editors. Practical Nuclear Medicine. 3rd Edition, London: Springer; 2005.
  • [4] Zeng GL. Image reconstruction – a tutorial. Comput Med Imag Grap 2001; 25: 97–103.
  • [5] Cherry SR, Sorenson JA, Phelps ME. Physics in Nuclear Medicine, 3rd Edition, Philadelphia: Saunders; 2003.
  • [6] Wernick MN, Aarsvold J, editors. Emission Tomography: The Fundamentals of SPECT and PET. 1st Edition, San Diego: Elsevier; 2004.
  • [7] Harrison C. Single photon emission computed tomography and its use in neuroimaging: a review for optometrists. Clin Eye Vis Care 1999; 11: 139–47.
  • [8] Zeng GL. Image reconstruction: a tutorial. Comput Med Imag Grap 2001; 25: 97–103.
  • [9] Kunyansky LA. A new SPECT reconstruction algorithm based on the Novikov explicit inversion formula. Inverse Probl 2001; 17: 293–306.
  • [10] Smith MF, Floyd CE, Jaszczak RJ, Coleman RE. Reconstruction of SPECT images using generalized matrix inverses. IEEE T Med Imaging 1992; 11(2): 165–75.
  • [11] Laerea KV, Koole M, Lemahieu I, Dierckx R. Image filtering in single-photon emission computed tomography: principles and applications. Comput Med Imag Grap 2001; 25: 127–33.
  • [12] Germano G. Technical aspects of myocardial SPECT imaging. J Nucl Med 2001; 42(10): 1499–507.
  • [13] Groch MW, Erwin WD. Single-photon emission computed tomography in the year 2001: instrumentation and quality control. J Nucl Med 2001; 29: 9–15.
  • [14] IAEA. Human Health Series No. 6: Quality assurance for SPECT systems. Vienna: International Atomic Energy Agency; 2009.
  • [15] Cerqueira MD, Matsuoka D, Ritchie JL, Harp GD. The influence of collimators on SPECT center of rotation measurements: artifact generation and acceptance testing. J Nucl Med 1988; 29: 1393–7.
Typ dokumentu
Bibliografia
Identyfikatory
Identyfikator YADDA
bwmeta1.element.baztech-c406b773-c945-4f05-a6ca-e5c487aa0e6a
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.