2004 | Vol. 24, nr 2 | 105-114
Tytuł artykułu

The simple method determining accuracy of the CMM software by means of virtual standard

Autorzy
Wybrane pełne teksty z tego czasopisma
Warianty tytułu
PL
Prosta metoda określenia dokładności oprogramowań WMP za pomocą wirtualnego wzorca
Języki publikacji
EN
Abstrakty
EN
In the article, the simple test that helps to determine the accuracy of CMM's software based on virtual 3D standard (regular pyramide) has been presented. The measure of the test's accuracy is the degree of the deformation of ideal pyramide. Computer and its software were considered as a „black box". Initial data were input through the keyboard to the computer co-operating with CMM. This way only error of the software and the computer itself affected the results, excluding the errors of the measuring machine. This way there has been proved the existence of large software errors that may even exceed uncertainty of the CMM.
PL
W artykule przedstawiono prosty test dokładności oprogramowania współrzędnościowej maszyny pomiarowej (WMP) oparty na wirtualnym wzorcu 3D (ostrosłup o podstawie czworokątnej). Miarą dokładności testu jest stopień deformacji prawidłowego ostrosłupa. Komputer i jego oprogramowanie traktowano jako „czarną skrzynkę". Dane wejściowe wprowadzono za pomocą klawiatury do komputera współpracującego z maszyną pomiarową. W ten sposób uzyskano końcowy wynik obarczony jedynie błędami samego oprogramowania i komputera, bez błędów maszyny pomiarowej. Wykazano możliwość powstania dużych błędów związanych z oprogramowaniem, które mogą być większe niż niepewność WMP.
Wydawca

Rocznik
Strony
105-114
Opis fizyczny
Bibliogr. 14 poz.
Twórcy
  • Institute of Mechanical Technology, Poznan University of Technology
Bibliografia
  • [1] Barker R.M., Harris P.M., Parkin G.I., Software support for metrology best practice guide No. 7: Development and testing of spreadsheet applications, National Physical Laboratory 2000.
  • [2] Brinkley D., Guide to the development of software for metrology SSfM software support for metrology best practice guide No. 3: Guidance on developing software for metrology, Logica (UK).
  • [3] Cox M.G., Forbes A.B., Harris P.M., Software support for metrology best practice guide No. 4: Discrete Modelling, 2002.
  • [4] Cox M.G., Harris P.M., Guidelines to help users: Select and use software for their metrology applications, National Physical Laboratory, Report CMSC 04/00 2000.
  • [5] DIN 32880 teil 1 (1986) Koordinatenmesstechnik Geometrische Grundlagen und Begrife.
  • [6] Emmet L., Froome P., Software support for metrology best practice guide No. 2: The development of virtual instruments, Adelard and Crown Copyright 1999.
  • [7] Internet: http://www.euroncap.com/.
  • [8] ISO 10360-6:2001(E) Geometrical product specifications (GPS) - Acceptance test and reverification test for coordinate measuring machines (CMM), Part 6: Estimation of errors in computing Gaussian associated features.
  • [9] Lin Y.-J., Damodharan K., Shakarji C., Standardised reference data sets generation for coordinate measuring machine (CMM) software assessment, The International Journal of Advanced Manufacturing Technology, 2001, No. 18, p. 819-830.
  • [10] Swornowski P., Wyznaczanie dokładności oprogramowań WMP przy pomocy korpusu 3D, in: IX Konferencja Naukowo-Techniczna "Metrologia w technikach wytwarzania maszyn", Częstochowa 2001, p. 535-540.
  • [11] Swornowski P., The errors of the CMM software algorithms shown on the example of the length measurement, Archives of Mechanical Technology and Automation, 2002, vol. 22, nr 2, p. 73-78.
  • [12] Swornowski P., The optimum number and distribution of measuring points for the circle with the shape deviation, Archives of Mechanical Technology and Automation, 2002, vol. 22, nr 2, p. 79-86.
  • [13] Waldele F., Bittner B., Dreschner R., Elligsen R., Test von Software fur Koordinatenmessgerate, VDI-Z, 1991, 12, p. 57-59.
  • [14] Wichmann B., Measurement good practice guide No. 5: Software in scientific instruments, National Physical Laboratory, Crown Copyright 1997.