EN
This paper describes a new contactless conductivity detector, whose electrodes are constructed of microchannels filled with solution of KCl - called pseudoelectrodes. The lab-on-a-chip microdevice was fabricated in poly(dimethylsiloxane) PDMS, using a moulding technique. The mould was made from a dry negative photoresist with a thickness of 50 µm. During the tests, the dimension and arrangement of pseudoelectrodes' microchannels were evaluated. The analyte was pumped into the microchannel using a syringe pump with a flow rate of 50 µL/min. Reproducible changes of the signal were obtained.