This work proposes a systematic assessment of measuring type A uncertainty (caused by random errors) used in RF power sensor calibration. To reduce A type uncertainty, several successive measurements are repeated. The uncertainty arises from repeatability errors in connectors caused by changes in their electrical properties during repeated mating. The suitability of the METAS UncLib software was analysed and we concluded that software should be developed to take into account the shape of probability density function (PDF) using a Monte Carlo method (MCM), which was lacking in METAS UncLib. The self-developed software was then tested on an example taken from the literature and the superiority of the MCM over the analytical method (GUM) was confirmed. During the calibration of the RF sensor using a vector network analyzer (VNA), a series of repeated measurements were performed and, after applying our MCM software, it was found that the measurement uncertainties calculated by the MCM method were several times larger than those by the GUM. The reason for this was that the correlation between the measured input quantities was not taken into account. When this was done using a covariance matrix and assuming a normal PDF of the input quantities, the results obtained with the GUM and the MCM converged. Our main objective was to investigate the influence of the PDF shape of the input measurement samples on the measurement uncertainty. Taking more than a dozen measurements is too costly, on the other hand, the small sample size prevents a reliable determination of the PDF shape. Finally, to overcome this inconvenience, we have developed a special method that uses the histograms of standardized input data taken at all measurement frequencies under fixed conditions without disconnecting the connectors, to increasing the total number of results which were needed to create the PDF histograms of input quantities.
Calibration of RF power sensors is crucial issue in RF power measurements. Many calibration laboratories use the direct comparison transfer system with a signal generator and a power splitter. Increasing performance of modern vector network analyzers makes it possible to perform a power sensor calibration with acceptable uncertainties. The main advantage when using a VNA is a simple measurement setup with a wide frequency range (up to 50 GHz, limited only by the VNA and the standard power sensor), where all of required components, i.e. signal generator, a directional coupler and a reference power indicator are built in the VNA technology. This paper reports performing a VNA-based RF power sensors calibration for 10 MHz – 18 GHz band, carried out in the Laboratory of Electric, Electronic and Optoelectronic Metrology at the National Institute of Telecommunications in Warsaw, Poland. In order to validate the proposed solution two of power sensors were calibrated at a reference laboratory. The validation consisted of two steps. At first, one of those characterized power sensors was calibrated at our laboratory in direct comparison transfer system. Finally, the results obtained from the VNA-based system were compared with the previously obtained ones.
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.