Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników

Znaleziono wyników: 1

Liczba wyników na stronie
first rewind previous Strona / 1 next fast forward last
Wyniki wyszukiwania
help Sortuj według:

help Ogranicz wyniki do:
first rewind previous Strona / 1 next fast forward last
EN
The aim of the work was measurement and testing of implantable cardioverter - defibrillator (ICDs) device. There was realised electronic measuring circuit which substitutes human body tissue during the device accuracy testing. This circuit enables to measure defibrillation shock discharge in the whole energy range (0.1- 41J). The presented measuring and testing procedures can easy check-up precision of parameters and functionality of embedded electronic device circuit’s part of the ICDs. Presented suggestion and results of this work can be implemented as verification procedure in various types of ICDs and also pacemakers before their implantation as well.
PL
Celem pracy było zbadanie właściwości elektrycznych rozrusznika serca. Zaproponowano procedure testowania tego typu urzadzeń.
first rewind previous Strona / 1 next fast forward last
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.