This article is about realization of the laboratory stand used for measuring parameters of LCD displays. The stand together with additional accessories (spectrophotometer and measuring probes) as well as software allow to measure a wide range of parameters (colour gamut, response time, contrast coefficient and its irregularity, luminance of black colour and its irregularity, luminance of white colour and its irregularity, changes of luminance depending on the viewing angle). This paper also show method and results of measurements of differences in contrast ratio and luminance uniformity between particular pieces of the same model of a display screen (research has been carried out on 5 pieces of the same model of a display). The presented results show that the measured value of the viewing angle may significantly differ from the value provided by producers of displays.