The construction and the general performance characteristics of thebaine-selective PVC membrane electrode based on thebaine-reinecke ion-pair complex was described. The electrode exhibited a rapid and Nernstian response (56.7 mV decade') to thebaine hydrochlo-ride concentration ranging from 1 x 10(-6)to I 10(-3)mol dm"1 within pH 1.0-5.0, The lower detection limit was 7.7 x 10(-7)' mol dm(-3). Interfering effects of some foreign substances were reported. The proposed electrode was applied for the direct assay of thebaine hydrochloride content in injections by means of the standard addition method. The results obtained were compared with the determination made by use of high performance liquid chromatography.
PL
Opisano budowę i zbadano charakterystykę elektrody membranowej selektywnej względem tebainy, która zawiera parę je nową: tebaina-sól Reinecke'go w PVC. Elektroda wykazuje nernstowską odpowiedź (56.7 mV dekada(-1) względem chlorowodorku tebainy w zakresie stężeń l x l0(-6) x 10(-3) mol dni(-3) i w zakresie pH 1.0-5.0. Granica wykrywalności wynosi 7.7 x 10(-7) mol dnr'. Wskazano na efekty interferencyjne ze strony niektórych substancji. Elektrodę zasotoswano do bezpośredniego oznaczania chlorowodorku tebainy w zastrzykach, za pomocą metody dodatków. Uzyskane wyniki porównano z wynikami uzyskanymi za pomocą wysokosprawnej chromatografii cieczowej.