Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników

Znaleziono wyników: 1

Liczba wyników na stronie
first rewind previous Strona / 1 next fast forward last
Wyniki wyszukiwania
Wyszukiwano:
w słowach kluczowych:  stripping chronopotentiometry
help Sortuj według:

help Ogranicz wyniki do:
first rewind previous Strona / 1 next fast forward last
EN
A flow-through stripping chronopotentiometry has been used for calibrationless determination of lead in wastewater samples. Lead was deposited on glassy carbon porous electrode, previously coated with Nafion® and mercury, applying a potentiostatic mode, and then stripped applying a constant current. The procedure was validated and applied to the analysis of wastewater samples from electroplating plant. Accuracy of the results was confirmed by graphite furnace atomic absorption spectrometry. LOD and LOQ for 5 mL-in-volume sample were 0.07 μg L-1 and 0-22 &mu.g L-1, respectively. Linear response was obtained for the analyte concentration range 0.5-2000 μ L-1for sample volumes from 0.5 to 5.0 mL. Repeatability andreproducibility were 1.1 % and 3.7%, respectively.
PL
Chronopotencjometrię w przepływie zastosowano do bezkalibracyjnego oznaczania ołowiu w próbkach ścieków. Ołówzatężano na porowatej elektrodzie z węgla szklistego uprzednio pokrytej warstwą Nafionu i rtęci. Metodę zwalidowano i zastosowano do analizy próbek ścieków z elektrolizerni. Uzyskane wyniki porównano z wynikami otrzymanymi metodą atomowej absorpcji z piecem grafitowym. Granice wykrywalności i oznaczalności, w przypadku próbek o objętości 5 mL, wynosiły odpowiednio 0.07 μg L-1. Sygnał był liniowy w funkcji stężenia oiowiu w zakresie 0.5-2000L-1, w przypadku próbek o objętości od 0,5 do 5,0 mL. Powtarzalność i odtwarzalność wynosiły odpowiednio 1,1 i 3,7%.
first rewind previous Strona / 1 next fast forward last
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.