Content available remote Miniature ion-selective electrode for flow measurements
EN
The paper presents the design and tests of a miniature ion-selective electrode for applications in flow measurements. A versatile head for sensor mounting was used to test the designed sensors. An ion selective membrane was fabricated at the end of a special plug mounted in the head. The sensors were tested in an automated measuring system consisting of an ion meter, a peristaltic pump, a burette and a personal computer. The system is controlled by special application software developed under the LabView environment.
PL
W pracy opisano nową konstrukcję miniaturowej elektrody jonoselektywnej dostosowanej do pomiarów przepływowych. Elektroda została wykonana w taki sposób, aby możliwe było jej zastosowanie w stosowanej głowicy pomiarowej. Membrana jonoselektywna została wytworzona bezpośrednio w specjalnym korpusie wtyczki montowanej w głowicy. Wykonane czujniki testowano w zautomatyzowanym systemie składającym się z jonometru, pompy perystaltycznej, biurety oraz komputera. Praca systemu była nadzorowana przez aplikację wykonaną w środowisku LabYiew.