The article presents a calibration method for a measuring circuit with a laser triangulation distance sensor. The authors describe factors influencing the accuracy of measurement using a laser triangulation sensor. The computer software supporting calibration procedure was developed, and the control-measuring software of the penetrometer was modified. The method is the result of the process of improvement of the penetrometer. The application of the developed penetrometer calibration method helps to achieve assumed measuring accuracy of ±2 units of penetration.
PL
W artykule przedstawiono metodę kalibracji toru pomiarowego z laserowym triangulacyjnym czujnikiem odległości. Omówiono czynniki wpływające na dokładność pomiaru za pomocą triangulacyjnego czujnika laserowego. Opracowano oprogramowanie komputerowe wspomagające procedurę kalibracji oraz zmodyfikowano oprogramowanie pomiarowo-sterujące penetrometru. Metoda jest wynikiem procesu doskonalenia penetrometru. W wyniku aplikacji opracowanej metody kalibracji penetrometru uzyskano zakładaną dokładność pomiaru ±2 jednostki penetracji.