Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników

Znaleziono wyników: 1

Liczba wyników na stronie
first rewind previous Strona / 1 next fast forward last
Wyniki wyszukiwania
Wyszukiwano:
w słowach kluczowych:  code AlfaMC
help Sortuj według:

help Ogranicz wyniki do:
first rewind previous Strona / 1 next fast forward last
EN
Measurements of α-particle sources require corrections to the counting rate due to scattering and self- -absorption in the source and the backing material. In this study, we describe a simple procedure to estimate these corrections using the new Monte Carlo code AlfaMC to consider the effects of scattering and self-absorption conjointly, and so to determine the activity of α emitters. The procedure proposed was applied to 235UO2 sources deposited on highly polished platinum backings. In general, the dependence of the efficiency with source thickness was in good agreement with a simple model considering a linear and a hyperbolic behavior for thin and thick sources, respectively, although significant deviations from this model were found for very thin sources. For these very thin sources, the Monte Carlo simulation revealed to be as a required method in the primary calibration of α-particle sources. The efficiency results obtained by simulation with AlfaMC were in agreement with available efficiency data.
first rewind previous Strona / 1 next fast forward last
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.