Znaleziono wyników: 9

Liczba wyników na stronie
first rewind previous Strona / 1 next fast forward last
Wyniki wyszukiwania
Wyszukiwano:
w słowach kluczowych:  badania tarciowo-zużyciowe
help Sortuj według:

help Ogranicz wyniki do:
first rewind previous Strona / 1 next fast forward last
EN
The polymeric components of endoprostheses in in vivo conditions are subjected to the oxidation process (ageing). The paper presents the results of friction and wear testing performed with use of a knee-joint simulator. The polymeric parts were machined from UHMWPE and UHMWPE modified with the addition of vitamin E. The calorimetric analysis and tribological testing results confirm the negative effect of accelerated ageing in oxygen atmosphere on the tribological properties of UHMWPE as well as the antioxidative properties of vitamin E.
PL
Polimerowe komponenty endoprotez w warunkach in vivo ulegają procesowi starzenia (utleniania). W artykule przedstawiono wyniki badań tarciowo-zużyciowych endoprotez stawu kolanowego. Komponenty polimerowe wykonano z UHMWPE i UHMWPE modyfikowanego witaminą E. Wyniki badań kalotometrycznych i tribologicznych potwierdziły negatywny wpływ procesu przyspieszonego starzenia w atmosferze tlenu na właściwości strukturalne i tribologiczne UHMWPE oraz antyoksydacyjne właściwości witaminy E.
PL
Rosnące wśród pacjentów zapotrzebowanie na endoprotezy stawów, związane m.in. ze starzeniem się populacji, wymaga od współczesnej medycyny ciągłego postępu i udoskonalania stosowanych w implantologii materiałów. W pracy przedstawiono wyniki wstępnych badań tarciowo-zużyciowych materiałów wybranych spośród powszechnie stosowanych w endoprotezoplastyce stawu kolanowego. Badania przeprowadzono z zastosowaniem testera T-17 pracującego w skojarzeniu badawczym typu trzpień–płytka (block–on–disk), przeznaczonego do wyznaczania charakterystyk tribologicznych skojarzeń materiałowych współpracujących ślizgowo w ruchu posuwisto- -zwrotnym, ze szczególnym uwzględnieniem materiałów przeznaczonych na elementy protez stawów człowieka, np. polimerów. Wyznaczeniu charakterystyk tribologicznych oraz określeniu zużycia liniowego poddano trzpienie wykonane z polietylenu o ultrawysokiej masie cząsteczkowej (UHMWPE) oraz polietylenu UHMWPE modyfikowanego witaminą E. Materiał przeciwpróbki (płytki) stanowił stop CoCrMo. Badania prowadzono przez milion cykli w środowisku surowicy bydlęcej o temperaturze 37°C±2°C. Częstotliwość ruchu wynosiła 1 Hz±0,1 Hz. W celu określenia mechanizmu zużycia, powierzchnie trące badanych materiałów po testach poddano obserwacjom przy pomocy skaningowej mikroskopii elektronowej. Analiza uzyskanych wyników badań tribologicznych pozwoliła wskazać spośród badanych próbek korzystniejsze skojarzenie materiałowe, którym była para trąca: modyfikowany witaminą E polietylen UHMWPE–stop CoCrMo. Dla powyższej pary trącej odnotowano nieznacznie większe wartości współczynnika tarcia, ale jednocześnie mniejsze zużycie liniowe. Natomiast na podstawie obserwacji morfologii powierzchni określono główny mechanizm zużycia, którym było zużycie ścierno-adhezyjne.
EN
The growing demand among patients for endoprosthesis joints associated, among others, with an aging population, requires from modern medicine continuous progress and improvement of materials used in implantology. The paper presents the preliminary results of friction-wear tests of materials chosen from among those commonly used in knee endoprostheses. The tests were conducted using a T-17 tester working in a block-on-disk tribosystem designated to determine the tribological characteristics of material combinations articulating in a reciprocating motion with particular reference to the materials used for elements of human joint prostheses, e.g. polymers. The tribological characteristics were determined and the linear wear was identified for pins made of ultrahigh molecular weight polyethylene (UHMWPE) and UHMWPE modified with vitamin E. The counter specimen materials (discs) consisted of a CoCrMo alloy. The study was conducted throughout a million cycles in an environment of bovine serum at a temperature of 37°C±2°C. The frequency of movement was 1 Hz±0.1 Hz. In order to determine the wear mechanism, the friction surfaces of the tested materials were subjected to observation using scanning electron microscopy after testing. Analysis of the tribological test results enabled the authors to indicate from among the tested samples the most advantageous material combination, which was the vitamin E modified UHMWPE– CoCrMo alloy friction pair. For the above friction pair a slightly higher friction coefficient was recorded, but also at the same time reduced linear wear. In contrast, based on the surface morphology observations, the main wear mechanism was defined which was abrasive-adhesive wear.
PL
Stosowanie bezniklowej stali austenitycznej 316L na implanty układu kostno-stawowego wiąże się z wieloma problemami eksploatacyjnymi. Jednym ze sposobów radzenia sobie z tymi ograniczeniami może być opracowanie nowych materiałów, w tym kompozytów, z zastosowaniem metalurgii proszków. W artykule przedstawiono wyniki badań tribologicznych spieków 316L i 316LHAp (hydroksyapatyt, HAp) otrzymanych metodą iskrowego spiekania plazmowego z zastosowaniem różnej temperatury spiekania. Dla stali 316L zwiększenie temperatury spiekania z 1000 do 1100°C skutkowało pogorszeniem odporności na zużycie. Niemniej w porównaniu z litą stalą 316L odporność na zużycie tych materiałów była znacząco większa. Dla kompozytów 316L-HAp zwiększenie temperatury spiekania skutkowało polepszeniem odporności na zużycie. Zużycie kompozytu spiekanego w niższej temperaturze było porównywalne ze zużyciem materiału referencyjnego.
EN
The use of 316L stainless steel in osteoarticular system implants is associated with many operational problems. One of the methods that deal with these restrictions may be the development of new materials, including composites, using powder metallurgy. This paper presents the results obtained during tribological tests of 316L and 316L-HAp sintered materials produced with the use of the spark plasma sintering method applying various sintering temperature. The increase of the sintering temperature from 1000 to 1100°C resulted in the deterioration of the wear resistance of 316L. However, compared to solid 316L, the wear resistance of these materials was several dozen times higher. The increase of the sintering temperature resulted in the improvement of the wear resistance of 316L-HAp. The wear resistance of the composites that had been sintered at lower temperatures was comparable to the reference material.
PL
Stop Ti6Al4V ze względu na swoje właściwości znajduje zastosowanie nie tylko w technice, ale także w medycynie, np. na implanty porowate, jak i lite. W przypadku implantów litych, ważnym problemem są niezadowalające właściwości tribologiczne tego stopu. Celem ich poprawy stosuje się nowe techniki wytwarzania do których zalicza się metodę iskrowego spiekania plazmowego (spark plasma sintering – SPS). W artykule przedstawiono wyniki wstępnych badań tribologicznych spieków z proszku Ti6Al4V wytworzonych tą metodą. Spiekanie z wykorzystaniem urządzenia HP D 25-3 prowadzono w temperaturze 1000°C w czasie 5 min przy ciśnieniu prasowania wynoszącym odpowiednio 5, 25 i 50 MPa z szybkością nagrzewania wynoszącą odpowiednio 200, 300 i 400°C/min. Wytworzone spieki poddano pomiarom gęstości i twardości, obserwacjom mikrostruktury oraz testom tarciowo-zużyciowym z wykorzystaniem testera tribologicznego T-05 (skojarzenie badawcze blok–pierścień) przy obciążeniu skojarzenia badawczego siłą odpowiednio 120, 220 i 320 N. Testy realizowano ze stałą prędkością wynoszącą 0,36 m/s na drodze tarcia 1000 m. Przed-stawiono zależność zmiany współczynnika tarcia w funkcji drogi tarcia oraz określono zużycie masowe i objętościowe badanych materiałów. Przeprowadzono również obserwacje morfologii powierzchni trących spieków. Wykazano, że najmniejszym zużyciem tribologicznym odznaczają się spieki wytworzone przy ciśnieniu prasowania wynoszącym 50 MPa z szybkością nagrzewania wynoszącą 400°C/min, a głównym mechanizmem zużycia tribologicznego było zużycie ścierne. Na powierzchniach trących badanych próbek nie stwierdzono śladów zużycia adhezyjnego.
EN
The Ti6Al4V alloy due to its characteristics is applicable not only in technology but also in medicine, such as porous implants and solid ones. In the case of solid implants, a major problem is inadequate tribological properties of this alloy and in order to improve them, new manufacturing techniques which include the method of spark plasma sintering are applied. The article presents the preliminary tribological results of sinters from Ti6Al4V powder produced by the SPS method. Sintering using an HP D 25-3 was carried out at 1000°C over 5 min at a compaction pressure of respectively 5, 25 and 50 MPa at a heating rate of 200, 300 and 400 °C/min. The resulting sinters were measured for density and hard-ness, subjected to microstructure observations and friction and wear tests using the tribological tester T-05 (block–ring tribosystem) at tribosystem load forces of 120, 220 and 320 N. The tests were carried out at a constant speed of 0.36 m/s on a friction path of 1000 m. The dependence of the friction coefficient changes as a function of friction path as well as specified mass and volume wear of the tested materials is presented. Observations of the bearing surface morphology of the sinters were also conducted. It was shown that the sinters fabricated at a compaction pressure of 50 MPa and heating rate of 400°C/min demonstrated the least tribological wear, and the main tribological wear mechanism was abrasive wear. On the bearing surface of the tested samples there were no signs of adhesive wear.
EN
Purpose: The paper describes the process of designing, manufacturing and design verification of the intervertebral of a new structure of lumbar disc endoprosthesis – INOP/LSP.1101. Methods: Modern and noninvasive medical imagining techniques, make it possible to record results of tests in a digital form, which creates opportunities for further processing. Mimics Innovation Suite software generates three-dimensional virtual models reflecting the real shape and measurements of components of L4-L5 spinal motion segment. With the use of 3D Print technique, physical models of bone structures of the mobile segment of the spine as well as the INOP/LSP.1101 endoprosthesis model were generated. A simplified FEA analysis of stresses in the endoprosthesis was performed to evaluate the designed geometries and materials of the new structure. Results: The endoprosthesis prototype was made of Co28Cr6Mo alloy with the use of selective laser technology. The prototypes were subject to tribological verification with the use of the SBT-03.1 spine simulator. Conclusions: The structure of the endoprosthesis ensures a full reflection of its kinematics, full range of mobility of the motion segment in all anatomical planes as well as restoration of a normal height of the intervertebral space and curvature of the lordosis. The results of the tribological tests confirmed that SLM technology has the potential for production of the human bone and jointendoprostheses.
PL
W artykule przedstawiono wyniki badań tribologicznych pary trącej „głowa-panewka”, stanowiącej układ ruchowy endoprotezy stawu biodrowego, o średnicy ?28 mm. W badaniach uSyto głowę wykonaną ze stopu CoCrMo oraz panewkę UHMWPE z powłoką C-C wykonaną w Instytucie Fizyki Jądrowej PAN w Krakowie. Badania prowadzono w Laboratorium Inżynierii Powierzchni i Tribologii Instytutu Obróbki Plastycznej w Poznaniu z wykorzystaniem symulatora SBT-01. Przeprowadzono badania momentu tarcia, siły docisku, wartości temperatury styku oraz zużycia elementów endoprotezy o średnicy 28 mm i porównano z wynikami pary trącej: głowa CoCrMo oraz panewka UHMWPE bez powłoki uformowanej jonowo. Podano i omówiono wyniki badań.
EN
The article presents results of tribological tests for 28 mm head-cup friction couple, which is the motion system of the hip endoprosthesis. In the studies femoral head made of CoCrMo alloy and polyethylene acetabular cup with C-C coating made in Institute of Nuclear Physics of Polish Academy of Science were applied. Investigations were performed in Laboratory of Surface Engineering and Tribology in Metal Forming Institute in Poznan using SBT-01 simulator. Friction moment, contact force, joint temperature and o 28 mm friction elements wear were investigated. This results was compared to results of friction couple: CoCrMo femoral head and UHMWPE acetabular cup without ion formed coating. There are presented and discussed investigations results.
PL
Przedstawiono metodykę tribologicznych badań materiałów i elementów endoprotez stawu biodrowego. Badania prowadzone są na dwóch etapach. Etap pierwszy dotyczy oceny różnych skojarzeń materiałowych pod względem oporów ruchu oraz badania odporności na zużycie za pomocą testera T-05 w styku rozłożonym. Etap drugi obejmuje badania tribologiczne par trących wytypowanych na pierwszym etapie pracy na symulatorze SBT-01 do badań elementów endoprotez stawu biodrowego. Symulator SBT-01 jest prototypowym urządzeniem badawczym zaprojektowanym i zbudowanym w Laboratorium Inżynierii Powierzchni i Tribologii Instytutu Obróbki Plastycznej w Poznaniu. Badania obejmują również analizę topografii i stanu powierzchni przed i po testach za pomocą profilometru, mikroskopu optycznego i SEM. W artykule zaprezentowano wyniki badań panewki z UHMWPE w skojarzeniu z głową ze stopu CoCrMo. Wyznaczono krzywe momentu tarcia, obliczono pracę tarcia dla pojedynczych cykli oraz określono zużycie elementów endoprotezy. Wyznaczono także ciepło wydzielone w procesie tarcia.
EN
This study contains the methodology of materials and hip endoprosthesis component investigations. Studies are realised in two stages. The first stage concerns the evaluation of the behaviour of material connections due to friction forces and wear resistance using a T-05 tester in distributed contact. The second stage contains tribological investigations of material connections in the hip simulator SBT-01 chosen in the first stage. The simulator is a prototype-testing machine designed and built in the Laboratory of Surface Engineering and Tribology of Metal Forming Institute. Research was extended to the analysis of surface topography and condition using a profilograph, optical microscope, and SEM. An examination of UHMWPE cup and CoCrMo alloy head was done. The authors have performed friction moments curves and calculated the work of friction for single cycles. In addition, the study obtained data on the wear of components and heat evolved in the friction process.
PL
Całkowita wymiana stawu biodrowego jest obecnie wysoce skuteczną i efektywną metodą przywracania funkcjonowania tego stawu. Podstawową cechą, która powinna cechować sztuczne stawy jest ich stabilność. W artykule opisano wyniki badań wpływu wzajemnego ustawienia elementów endoprotezy stawu biodrowego na wybrane parametry procesu tarcia. Badaniom poddano dwie pary trące o różnych średnicach, wykonane z różnych materiałów. Eksperyment wykonano na zaprojektowanym i zbudowanym w Instytucie Obróbki Plastycznej w Poznaniu symulatorze do badań elementów endoprotez stawu biodrowego SBT-01. Praca wykonana została we współpracy z Kliniką Ortopedii Ogólnej, Onkologicznej i Traumatologii na Uniwersytecie Medycznym im. Karola Marcinkowskiego w Poznaniu.
EN
Total hip replacement is very effective method in case of human joint functionality restoring. The most important property of artificial joints is their functional stability. Results of influence of reciprocal endoprosthesis components setting on chosen friction parameters are presented. Two friction couples with different diameter made of different materials has been tested. Experiment has been performed using hip endoprosthesis simulator SBT-01 designed and constructed in Metal Forming Institute hip endoprosthesis simulator SBT-01. Work was carried out in cooperation with General Orthopaedics, Oncology and Traumatology Clinic at Karol Marcinkowski Poznan University of Medical Sciences.
PL
W pracy podano wyniki własnych badań nad utlenianiem stopu tytanu Ti6A14V, stosowanego w medycynie i wpływem modyfikacji warstwy wierzchniej tego stopu na własności trybologiczne. Badania przeprowadzono dla pary trącej Ti6A14V - polietylen, stosowanej w endoprotezach stawu biodrowego i endoprotezach stawu kolanowego.