Content available remote Oxidation and degradation of polyethylene cups in hip joint prostheses
EN
The properties of polyethylene cups of hip prostheses produced by Johnson & Johnson Orthopaedics, De Puy International Ltd. (Great Britain) and Aesculap AG & Co. KG (Tuttlingen, Germany), removed after various times of implantation due to their premature wearing out, were compared by means of the analysis of FT-IR spectra and determination of the hardness and viability of macrophages and osteoblasts. The oxidation of polyethylene was found to be a main reason of Aesculap cups' damages.
PL
Porównano zmiany następujące w polietylenowych panewkach stawu biodrowego dwóch producentów (symbole próbek C// i CAE) usuniętych po różnych czasach implantacji z powodu ich przedwczesnego zużycia. Przebieg utleniania zbadano metodą FT-IR na przykładzie próbek CAE. Na tej podstawie określono wartości wskaźnika utleniania (rys. 3); utlenianie stanowi przyczynę uszkodzeń panewek CAE W toku ich użytkowania. Oba rodzaje próbek poddano pomiarom twardości (rys. 4) oraz ocenie żywotności makrofagów i osteoblas-tów.