EN
A sensitive, stability-indicating reversed-phase high-performance liquid chromatography with diode array detection (HPLC–DAD) method has been developed for the determination of TBI-166 and its 10 kinds of related impurities. Chromatographic separation was achieved on a Kromasil ODS column (250 mm × 4.6 mm, 5 μm), with a gradient elution of the mobile phase system consisting of acetonitrile and 1% ammonium formate solution (with 0.2% formic acid). The flow rate was 1.0 mL/min, and the detection wavelength was set at 251 nm. The method was validated according to the International Conference on Harmonization (ICH) guidelines with respect to selectivity, linearity, limits, accuracy, precision, and robustness. The calibration curves were linear from LOQ to 150% of the specification limit of impurity with correlation coefficients not less than 0.999. The limits of quantitation were between 0.123 and 0.257 μg/mL. Accuracy for the related substances was estimated by the recovery ranged from 94.6% to 111.2%. The method was proved to be reliable for the determination of related substances in TBI-166 bulk drug, which is essential and important in the quality control.