Six radon laboratories, which perform indoor radon measurements to jointly investigate temporal variations of radon concentration in houses in a couple of regions in Poland, participated in the intercomparison exercise. There are involved three commercially available Hungarian RadoSys systems and four own laboratory methods. All of them are based on the etched-track CR-39 detectors. The intercomparison was conducted in a calibration laboratory of the Central Laboratory for Radiological Protection (CLOR), Warsaw, Poland, accredited by the Polish Center for Accreditation (previously AP 101, at present AP 057). Comparison measurements were performed during three expositions in the CLOR’s radon calibration chamber under controlled normal climatic conditions and steady radon concentration. The results were referred to the value of the radon concentration determined by AlphaGUARD monitor traceable to the primary 222Rn standard in Physikalisch-Technische Bundesanstalt (PTB), Braunchweig, Germany. The mean performance ratio defined as a ratio of the reference to participant’s result range from 0.88 to 1.31.