Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników

Znaleziono wyników: 1

Liczba wyników na stronie
first rewind previous Strona / 1 next fast forward last
Wyniki wyszukiwania
Wyszukiwano:
w słowach kluczowych:  Doppler ultrasound method
help Sortuj według:

help Ogranicz wyniki do:
first rewind previous Strona / 1 next fast forward last
EN
This work was aimed at assessment of usability of Doppler ultrasound method for the analysis of fetal heart rate variability at a level of single heart beats. The purpose of this study was to check if today fetal monitors provide the signal of accuracy comparable with an electrocardiography method. The virtual instrumentation software for measurement system and processing of acquired signals was implemented using LABVIEW environment. The results obtained reveal that accuracy of today monitors which use Doppler ultrasound technique is sufficient for visual analysis of the FHR traces. However, for the computerized analysis at a beat-to-beat level, this accuracy is below the acceptable value.
first rewind previous Strona / 1 next fast forward last
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.