Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników

Znaleziono wyników: 1

Liczba wyników na stronie
first rewind previous Strona / 1 next fast forward last
Wyniki wyszukiwania
Wyszukiwano:
w słowach kluczowych:  106Ru ophthalmic applicators
help Sortuj według:

help Ogranicz wyniki do:
first rewind previous Strona / 1 next fast forward last
EN
A gauge was developed for determining the dose rate distribution and surface activity of ophthalmic brachytherapy applicators, particularly for 106Ru applicators. A plastic fi 2×2 mm scintillator is used as the radiation detector, featuring a high pulse count rate, which results in a low 0.5% random error, due to good counting statistics. Automatic gain control of the photomultiplier tube (PMT) is achieved using a LED as the reference light source. The PMT operates in pulse mode. Long term gain variation due to fatigue of the PMT or ambient temperature variation is thus compensated for. The count rate error due to inaccurate setting of the high voltage supply of the PMT is 0.4%, and the instability error over 7 hours of continuous operation does not exceed 1-2%, peak-to-peak.
first rewind previous Strona / 1 next fast forward last
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.