EN
Square-wave anodic stripping voltammetry was applied to the simultaneous determination of Tl and Cd in wastewater samples with extremely high contents of the studied elements using a bismuth film electrode. The stripping signals were recorded in the supporting electrolyte containing acetate buffer (pH 4.5) and Bi3++ ions without removal of oxygen. It was observed that the stripping peak of Tl was insufficiently separated from the peak of Cd; however, quantification of the metals was possible when their concentrations were comparable. To verify the obtained results, the ICP-MS method was applied.
PL
Do jednoczesnego oznaczenia Tl i Cd w próbkach wód odpadowych charakteryzujących się wysokim stężeniem badanych metali zastosowano anodową woltamperometrie stripingową fali prostokątnej na blonkowej elektrodzie bizmutowej. Sygnały rejestrowano bez odtleniania w buforze octanowym (pH 4.5) w obecności jonów bizmutu(III). Pomimo zbliżonych potencjałów pików Tl i Cd można było ilościowo oznaczyć zawartość obu analitów w próbce, jeśli występowały w porównywalnych stężeniach. Otrzymane wyniki porównano z wynikami uzyskanymi metodą ICP-MS.