Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników

Znaleziono wyników: 1

Liczba wyników na stronie
first rewind previous Strona / 1 next fast forward last
Wyniki wyszukiwania
help Sortuj według:

help Ogranicz wyniki do:
first rewind previous Strona / 1 next fast forward last
EN
New, sensitive, rapid, cost-effective, and validated stability-indicating thin layer chromatographic (TLC) method coupled with fluorescence (FL) detection was developed for the quantitative analysis of celecoxib (CEL) and amlodipine besylate (AMLO) in their laboratory prepared binary mixture using the non-fluorescent TLC silica gel 60 plates. Ethyl acetate: diethylamine: 1-propanol (9:1:0.2, V/V) was used as a developing system. The retention factor (Rf) for each drug was 0.80 ± 0.03 and 0.44 ± 0.01 for CEL and AMLO, respectively. The plates were excited at 264 nm for the simultaneous FL measurement of CEL and AMLO, the calibration curves were linear over a concentration ranges of 30.0–300.0 ng/band and 15.0–150.0 ng/band with mean percentage recoveries of 99.80 ± 0.85 and 99.80 ± 0.77 For CEL and AMLO, respectively. The developed method was applied for the stability studies of the cited drugs in their laboratory prepared binary mixture and the forced degradation products were determined when present in presence of the pure drugs so the method can be considered as a stability-indicating one and it was validated as per ICH guidelines and proved to be accurate and precise.
first rewind previous Strona / 1 next fast forward last