EN
This paper presents the concept, design and experimental results of tests, in laboratory conditions, of a 2-port, space-saving pneumatic probe. The main features of the probe are a result of restrictions imposed by the functioning environment: a limited space for probe installation, the possibility of processing only two output signals and requirements for simplicity and reliability of the probe. A cylindrical shape of the probe tip is proposed as a general concept, similar to the classic, cylindrical, 3-port probe. The main difference arises in using only two pressure signals: one from the overpressure zone on the front side of the cylinder and another, from the underpressure zone on the back side. After performing an appropriate calibration procedure, it is possible to measure the flow velocity and correct the pressure difference obtained by means of a velocity coefficient kv. This paper also presents an analysis of the kv coefficient uncertainty to evaluate the quality of measurements.
PL
W artykule poddano analizie problematykę doboru funkcji aproksymujących przebiegi prostych i odwrotnych charakterystyk statycznych termoanemometrów CTA. Oprócz wykorzystywanych zwykle, typowych wskaźników dopasowania krzywej jak np. współczynnik determinacji, jako alternatywne kryterium wyboru postaci funkcji aproksymującej przyjęto wskaźnik, opisujący położenie krzywej aproksymującej w stosunku do pasma niepewności wynikającego z dokładności zastosowanego przyrządu wzorcowego. Analizy dokonano na podstawie wzorcowań przeprowadzonych z wykorzystaniem wielootworkowej sondy spiętrzającej (tzw annubary) jako przepływomierza wzorcowego.
EN
The questions of fitting curve choice, for the approximation of direct and indirect static characteristics of CTA anemometers were analyzed in the paper. Besides commonly used approximation quality criteria as the coefficient of determination, a parameter, describing relative position of fitting curve in relation to uncertainty zone, resulting from the accuracy of applied measurement device was applied. The analysis was performed basing on the results of calibrations, made with use of multihole impact probe (annubar) as a standard device.
3
Content available remote Ocena przydatności sond typu S w pomiarach prędkości przepływu
PL
W artykule przedstawiono wyniki wzorcowań ośmiu sond pneumatycznych typu S i dokonano oceny ich właściwości metrologicznych. Dla przebadanych sond stwierdzono stosunkowo duży rozrzut wartości współczynnika pozwalającego związać mierzoną różnicę ciśnień z wartością prędkości przepływu. Uzyskane różnice wskazują na brak możliwości przeniesienia wyników wzorcowań pomiędzy różnymi sondami tego typu - niezbędne jest wzorcowanie każdego z egzemplarzy osobno. Dokonano także analizy możliwych przyczyn uzyskanej rozbieżności parametrów poszczególnych sond.
EN
The paper presents the calibration results for eight different S-type pneumatic probes as well as the evaluation of their metrological features. A large spread of the coefficient - allowing to relate the measured pressure difference to the flow velocity - was found in evaluating the probes. The differences found indicate that calibration results are not transferrable between different probes of this type and that probes need to be calibrated individually. Potential reasons for the spread of the parameters obtained for different probes were also examined.
4
Content available remote Wybrane zagadnienia aproksymacji charakterystyk statycznych termoanemometrów CTA
PL
W artykule przeanalizowano możliwości aproksymacji odwrotnych charakterystyk termoanemometrów stałotemperaturowych (CTA) za pomocą różnych funkcji. Porównano wyniki aproksymacji z wykorzystaniem funkcji wielomianowych (3-go i 4-go stopnia), wykładniczych (eksponencjalnej, Stirliga oraz Gompertza) a także funkcji potęgowej. Wykorzystując wskaźniki jakości aproksymacji (takie jak współczynnik determinacji czy błąd średniokwadratowy) dokonano ilościowej oceny poszczególnych metod na bazie trzech różnych zestawów danych pochodzących z rzeczywistych wzorcowań termoanemometrów.
EN
Possibilities of approximation of inverse static characteristics of constant temperature hot-wire anemometers, by means of different functions were analyzed in the paper. The results of approximation with use of polynomials (3rd and 4th degree), exponential functions (exponential growth, Stirling and Gompertz function) as well as power function were presented and compared. Applying some coeffi cients describing the quality of approximation (like coefficient of determination or mean squared error), quantitative evaluation of particular methods was done, on the basis of three different data sets, originating from real hot wire calibrations.
PL
W pracy zaproponowano modyfikację metody stosowanej w pomiarach szybkozmiennych płynów stosowaną w technikach wytwarzania. Przedstawiono kluczowe cechy termoanemometrów CTA w kontekście pomiarów chwilowej szybkozmiennej prędkości przepływu w warunkach silnej zmienności temperatury czynnika. Przedstawiono przykładowe wyniki wzorcowania sondy termoanemometrycznej i na tej podstawie, zaproponowano procedurę budowy trójwymiarowej charakterystyki statycznej sondy termoanemometrycznej. Propozycję przedstawiono w trzech wariantach (trzy postacie równania powierzchni aproksymującej), z których zostało wybrane najlepsze rozwiązanie w oparciu o wskaźniki jakości aproksymacji wynikające ze statystyki matematycznej.
EN
A modification of the method used in instantaneous fluid measurements in production engineering is proposed. Key feature of CTA hot-wire anemometers under fast temperature changes are presented. Some results of the described hot-wire anemometer calibration were the basis of the proposed procedure of 3D static characteristics. The proposal is presented in tree versions, which have been selected according to the quality indicators based on mathematical statistics.
6
Content available remote Analiza możliwości zwiększania mocy tłokowych silników spalinowych
PL
W artykule przedstawiono analizę czynników wpływających na moc tłokowego silnika spalinowego. Wyprowadzono zależność pozwalającą wyrazić moc jednostkową silnika jako funkcję kilku zmiennych (parametrów), które poddano szczegółowej analizie w kolejnych podrozdziałach. Stwierdzono, iż wymierny wzrost mocy silnika o określonych wymiarach głównych można uzyskać głównie dzięki dwóm czynnikom: poprawie sprawności napełnienia oraz wzrostowi gęstości powietrza na wlocie do cylindrów. Szczególnie ten drugi sposób cechuje duża skuteczność, a dodatkowo pozwala on na kształtowanie charakterystyk zewnętrznych silnika w sposób korzystny z punktu widzenia sposobu jego użytkowania (np. właściwości trakcyjnych napędzanego pojazdu).
EN
The paper presents an analysis of parameters impacting the power of an IC engine. A relationship expressing the engine unit power as a function of some variables (parameters) had been found and those parameters had been thoroughly examined in successive subsections. It was found that - for a fixed engine size - a substantial increase of the engine power can be primarily attributed to two parameters: volumetric efficiency and air density at the engine inlet. The second parameter is particularly efficient in increasing the engine power and, additionally, it allows enhancing the shape of the engine characteristics from the point of view of its operating conditions (e.g. traction properties of a driven vehicle.)
PL
Artykuł przedstawia charakterystykę i źródło dostaw rud żelaza. Opisany jest wpływ udziału rud hematytowych i magnetytowych w mieszance spiekalniczej na wydajność taśmy spiekalniczej [1].
EN
The article presents the characteristics and sources of supply of iron ores. The influence of the share of hematite and magnetite ores in the sintering mix on the performance of the sinter strand is described [1].
EN
The phenomena occurring in a pulsating flow of gas through the straight pipe of a circular cross-section have been analyzed in this dissertation. Measurement methods that allow for determination of unsteady flow parameters (pressures, velocities and temperatures) for the pulse frequency up to /=200 Hz have been applied and improved. A method that enables to distinguish two components of the measured temperature signal, corresponding to static and dynamic temperature, has been proposed as well. In order to analyze the observed unsteady phenomena, a Bannister wave model that enables the pressure signal decomposition into the incoming and reflected wave has been adapted. Due to the employment of this model, the flow behavior in the types of systems under investigation could be explained to a great extent. A trial to adapt the classical acoustic theory to pulsating flow conditions has also been undertaken in the dissertation. The resonance frequencies of the systems under consideration have been determined with this method and a very good convergence of the calculation and measurement results has been obtained. The amplification coefficients for pressures and specific mass flow rates defined in the study reach their maximum values for the theoretically determined resonance frequencies. It has been also found that the pipe supplied with a pulsating flow can be treated dynamically as an object of the second order, which is confirmed by the amplitude and phase characteristics determined experimentally. To widen our knowledge on the character of flow parameter dynamics in the analyzed measurement cross-sections, unsteady temperature and specific mass flow rate fields have been measured for some selected conditions of excitation. It has been found that the pulsating flow alters the shape of the specific mass flow rate profile, leading to its uniformity. In the case of temperature, we encounter a very high degree of homogeneity of the unsteady fields recorded at the pipe inlet, which also indicates indirectly the homogeneity of static pressure fields. As regards practical applications of these investigation results, apart from typical cases of a flow through inlet and outlet pipes of diesel engines and piston compressors, they can be employed in all installations where a pulsating flow occurs and phenomena such as an increased pulsating amplitude or a reverse flow cause disturbances in their correct operation.
9
PL
W artykule opisano wyniki pomiarów rozkładów temperatury gazu w kanale o przekroju kołowym zasilanym przepływem pulsacyjnym. Przeprowadzono analizę własności dynamicznych zastosowanych przetworników temperatury oraz omówiono procedurę wykonania pomiarów, polegającą na sondowaniu badanego przekroju wzdłuż jego promienia. Przeanalizowano uzyskane wyniki dokonując próby skorelowania uzyskanych rozkładów temperatury z rozkładami innych parametrów przepływu (ciśnienia i prędkości przepływu). Stwierdzono istnienie niestacjonarnego rdzenia przepływu, w którym obserwuje się powtarzalność przebiegów temperatury na kolejnych badanych promieniach. W miarę zbliżania się do ścianki kanału ujawnia się wpływ wymiany ciepła przez ściankę. Wykazano także istotny wpływ dynamicznej składowej temperatury na rejestrowane przebiegi.
EN
The paper describes results of gas temperature distribution measurements in the circular pipe supplied with a pulsating flow. The analysis of dynamic properties of applied temperature sensors was conducted and the procedure of measurements was described as well. The temperature field survey was done by displacing the probe along the pipe radius. The results obtained were carefully analyzed and an attempt to find correlations between main flow parameters (pressure, temperature, velocity) was made. It has been found that the "flow core" exists in a large zone around the pipe axis. The repeatability of successive temperature plots, recorded at different radii, can be observed inside of it. While approaching the pipe wall, an influence of heat exchange through the pipe wall is manifested. It has been also shown that the dynamic temperature component can exert a strong influence on recorded temperature signals.
EN
The computational calculations employing the “x-t” model of a pulsating flow in the pipe and their experimental verification were described. The model, previously used for calculations of pressure signal variations in pneumatic transmission pipes, was adapted to the admission system of the turbocharger turbine. Appropriate boundary conditions were applied, using the experimental data provided by the measurements of unsteady flow parameters. The calculation results were compared to the experimental data showing a good agreement.
PL
W artykule opisano sposób wyznaczania niepewności pomiaru strumienia masy gazu w kanale przepływowym za pomocą wielootworkowych sond spiętrzających (tzw. Annubar probes). Pomiary realizowane były metodą pośrednią derywacyjną. Analizie poddano składowe niepewności związane z przetwarzaniem sygnałów w poszczególnych torach pomiarowych oraz ich wpływ na końcową niepewność pomiaru. W analizie wykorzystano zalecenia ISO odnośnie wyznaczania niepewności. Wykazano, iż końcowa niepewność pomiaru strumienia masy zależy od konfiguracji torów pomiarowych, przy czym kluczowa jest w tym przypadku dokładność pomiaru ciśnienia różnicowego. Przy odpowiednio dobranej rozdzielczości przyrządu mierzącego to ciśnienie można uzyskać znacznie niższe wartości niepewności niż deklarowane przez producenta.
EN
The paper describes a method of determination of the mass flow rate uncertainty in case of gas flow measurements executed with multihole impact probes (Annubar probes). Measurements were conducted by means of an indirect, derivative method. The uncertainty components connected with successive measuring chains were analyzed and their influence on the resultant combined uncertainty was examined as well. During the analysis, the recommendations of ISO standards as regards uncertainty expression were applied. It was shown that the resultant uncertainty of the mass flow rate depends on the configuration of successive measuring chains and the key element of the measuring process is the accuracy of differential pressure measurement. Applying a measuring device of appropriate resolution, it is possible to obtain lower mass flow rate uncertainties than those declared by the producer.
12
Content available remote Specific mass flow rate measurements in a pulsating flow of gas
EN
The method of measurement of the specific mass flow rate in a pulsating flow by means of the Constant Temperature Anemometer (CTA) has been described. Special attention has been paid to the probe calibration problem. Different forms of the CTA characteristic have been analyzed. Example measurement results have been presented for chosen pulse frequencies. Apart from the measurements executed in one single point representing each control section, a flow field survey has been done in order to determine the velocity profile under the conditions of a pulsating flow. Probes have been displaced radially with a small step to cover the range from the pipe axis to its wall. It has been found that in the large field around the pipe axis, successive velocity plots are similar as far as their shape and phase shift are considered. Pulsations are damped and mean velocity decreases rapidly only in direct proximity of the pipe wall. It has been also shown that the presence of pulsations makes the mean velocity profile more uniform than in the case of a steady flow. Measurements have been also performed for the case of a reverse flow occurring in the pipe at resonance frequencies.
PL
W artykule opisano wyniki pomiarów temperatury gazu w przepływie pulsacyjnym przez rurociąg, zrealizowane za pomocą cienkowłóknowego termometru stałoprądowego CCT w zakresie częstości wymuszenia 20÷200 Hz. Zarejestrowane sygnały poddano analizie, mającej na celu określenie warunków, w jakich konieczne jest uwzględnienie dynamicznej składowej temperatury, wynikającej z wyhamowania czynnika na powierzchni termometru. Zaproponowano metodę pozwalającą na wyodrębnienie z sygnału mierzonej temperatury składowych odpowiadających temperaturze statycznej i dynamicznej.
EN
The results of gas temperature measurements in the pipe under the conditions of the pulsating flow are described. The measurements were conducted with a thin wire CCT thermometer in the range of excitation frequencies of 20÷200 Hz. The acquired signals were analysed in order to define the conditions in which it is necessary to take into account the dynamic temperature component occurring as a result of velocity deceleration on the thermometer surface. The method of decomposition of the temperature signal allowing to separate its static and dynamic component is proposed.
PL
W pomiarach zmiennej w czasie temperatury płynącego gazu występują dwa kluczowe problemy: pierwszy związany jest z niedostateczną dynamiką stosowanych zwykle przetworników temperatury, drugi wynika z konieczności uwzględnienia dynamicznej składowej temperatury wynikającej z częściowego wyhamowania czynnika na powierzchni termometru. W artykule proponuje się sposób kompleksowego rozwiązania tego zagadnienia poprzez zastosowanie metody korekcji dwutorowej, której celem jest wyeliminowanie (lub wyraźne zredukowanie) błędu dynamicznego pomiaru temperatury a następnie rozłożenie sygnału temperatury na składowe odpowiadające temperaturze statycznej (wynikającej ze zmian parametrów termodynamicznych czynnika) oraz dynamicznej (związanej z prędkością przepływu gazu). W procesie tego rozkładu niezbędna jest znajomość charakterystyki tzw. współczynnika odzyskania energii kinetycznej określającego stopień przetworzenia energii kinetycznej gazu na ciepło rejestrowane jako wzrost temperatury czujnika. W artykule zaproponowano metodę wyznaczania takiej charakterystyki za pomocą sondy Prandtla o specjalnie zmodyfikowanej konstrukcji.
EN
Two key problems occur during measurements of unsteady temperature of the flowing gas: the first issue is related to insufficient dynamics of usually used temperature transducers, the second one follows from the necessity of taking into account the dynamical component of temperature that results from a partial impeding of the medium on the thermometer surface. A complex solution to this problem through an application of two-path correction method to eliminate (or reduce considerably) the dynamic error of the temperature measurement, and then a decomposition of the temperature signal into components corresponding to static temperature (resulting from changes in thermodynamic paramters of the medium) and dynamic temperature (related to gas flow velocity), is postulated. During the decomposition, it is necessary to know the characteristics of the so-called kinetic energy recovery coefficient that defines a degree of transformation of the gas kinetic energy into heat, recorded as an increase in the sensor temperature. In the paper, a method for determination of such characteristics with a specially modified Prandtl probe is discussed.
PL
W artykule opisano algorytm korekcji dwutorowej własności dynamicznych miniaturowych przetworników temperatury wykorzystującej sygnały z dwóch równoległych torów pomiarowych. Przedstawiono istotę metody oraz sposób jej implementacji w komputerowym systemie pomiarowym. Zaprezentowano również wyniki weryfikacyjne w odniesieniu do zmiennych w czasie sygnałów temperatury gazu, zarejestrowanych za pomocą miniaturowych czujników temperatury w przewodzie turbosprężarki zasilanym w sposób pulsacyjny.
EN
In the paper the identification algorithm of dynamic properties of temperature sensors by two chain correction method has been presented. Principles and practical applications of the correction method within computerized measuring system were discussed. Experimental results obtained for the time-varying gas temperature measurements performed using miniature temperature transducers in the turbocharger pipe supplied with pulsating gas flow have been given.
16
Content available remote Korekcja błędu dynamicznego miniaturowych przetworników temperatury gazu
PL
W artykule opisano podstawy korekcji własności dynamicznych miniaturowych przetworników temperatury oraz przedstawiono wyniki praktycznego jej zastosowania w odniesieniu do zmiennych w czasie sygnałów temperatury gazu, zarejestrowanych za pomocą miniaturowych czujników temperatury w przewodzie zasilanym w sposób pulsacyjny. Szczególny nacisk położono na metodę tzw. korekcji dwutorowej, wykorzystującej sygnały z dwóch równoległych torów pomiarowych. Porównano rezultaty otrzymane na podstawie tej metody oraz obliczeń modelowych, bazujących na uproszczonym modelu wymiany ciepła na spoinie termoelementu.
EN
Principles and practical applications of a dynamic properties correction of miniature temperature transducers are described in the paper. Experimental results concern the measurement of the time-varying gas temperature, which is read by the miniature temperature sensors placed in the pipe supplied with pulsating gas flow. Special attention was drawn to the dual-channel correction method based on the signals provided by two parallel measuring chains. Temperature sensor dynamic parameters obtained from this method were compared to the results of calculations based on the simplified model of heat exchange at the thermocouple junction.
EN
In the paper, the characteristic features of an unsteady flow in two systems: a high-volume pipe-receiver and a pipe-turbine are described. The tests were made under the conditions of a pulsating pipe supply in the range of frequencies 20-100 Hz. The following measurements of instantaneous values of unsteady flow parameters were made: pressure (piezoresistive transducers), temperature (Constant Current Thermometer) and velocity (Constant Temperature Anemometer). The applied measurement methods - owing to good dynamic properties of transducers - permitted us to gather relatively rich and complete information on flow phenomena in such systems. It was shown that the pipe-receiver system behaved as a resonator with a marked range of amplitude amplification and a distinct relationship between the resonance frequency and the pipe length. In this system, an effect of high pressure signal attenuation along the pipe and a reverse flow were observed as well. On the other hand, the pipe-turbine system is free of characteristic resonance frequencies and the pressure signal is transmitted along the pipe up to the turbine inlet section practically without any changes.
PL
W artykule opisano zjawiska przepływowe zaobserwowane w układach przewód-zbiornik oraz przewód-turbina zespołu ładującego przy pulsacyjnym charakterze zasilania tych układów. Wykazano, iż układ ze zbiornikiem zachowuje się jak rezonator, z wyraźnym obszarem wzmocnienia amplitud wszystkich badanych parametrów przepływu (ciśnienia, temperatury i prędkości). Przy stałej objętości zbiornika, położenie tego obszaru zależy od długości przewodu, co wskazuje na podobieństwo do rezonatora Helmholtza. Jednak przyjęcie modelu układu przepływowego w tej postaci nie daje zbieżności wartości częstości rezonansowych wyznaczanych z modelu oraz z pomiaru. Zbieżność tę, daje dopiero zastosowanie jako modelu układu Lutza, uwzględniającego odcinek przewodu za zbiornikiem. Charakterystyczną cechą układów ze zbiornikiem jest tłumienie sygnału ciśnienia wzdłuż długości przewodu, z kolei w układzie z turbiną, takie tłumienie jest pomijalne, sygnał w zasadzie bez zniekształceń jest przenoszony wzdłuż długości przewodu do przekroju wlotowego turbiny. W układzie tym nie występują również efekty rezonansowe. Analiza opisanych zjawisk przepływowych była możliwa dzięki zastosowaniu bardzo szybkich przetworników mierzonych parametrów przepływu.
PL
W artykule opisano dwie metody analizy niestacjonarnych sygnałów pochodzących z przetworników ciśnienia, temperatury i prędkości przepływu, zainstalowanych w prostym przewodzie zamkniętym zbiornikiem o dużej objętości. Zastosowano klasyczną metodę analizy sygnałów w oparciu o rozwinięcie w szereg trygonometryczny Fouriera oraz metodę polegającą na przedstawieniu sygnału ciśnienia jako superpozycji fal biegnącej i odbitej z wykorzystaniem modelu Bannistera. Zaprezentowano bogaty materiał badawczy i przeanalizowano przydatność obydwu tych metod do analizy tego typu sygnałów.
EN
In the paper two methods of unsteady signals analysis were presented. These signals provide from pressure, temperature and flow velocity transducers, mounted in the straight pipe closed at one end with use of a big volume tank. Two methods were applied: the classical method of analysis based on the trigonometric Fourier series expansion and the method based on pressure signal decomposition (expression of the pressure signal as a superposition of rightward and leftward pressure waves) with use of the Bannister model. Rich experimental material was presented and the suitability of both methods for the analysis of such signals was analysed
PL
W artykule opisano wyniki badań niestacjonarnego przepływu powietrza w przewodzie zasilanym w sposób pulsacyjny. Badania przeprowadzono dla dwóch różnych układów przepływowych, różniących się elementem zamykającym przewód. W pierwszym przypadku był to zbiornik o dużej objętości, w drugim turbina zespołu ładującego. Celem takiej konfiguracji stanowiska było zamodelowanie stałego (w przypadku zbiornika) oraz zmiennego (w przypadku turbiny) ciśnienia na wylocie z badanego przewodu. Zastosowano zaawansowane metody pomiarów chwilowych wartości parametrów przepływu (miniaturowe przetworniki piezorezystancyjne do pomiaru ciśnień, termometr stałoprądowy CCT do pomiaru temperatur oraz termoanemometr CTA do pomiarów prędkości przepływu). W efekcie przeprowadzonych pomiarów ujawniono odmienny charakter zjawisk przepływowych dla dwóch badanych typów układów. Wykazano także, iż dla pewnych częstości wymuszenia następuje wyraźny wzrost amplitudy pulsacji, a położenie jej maksimum jest funkcją długości przewodu. Pokazano także korelacje istniejące pomiędzy poszczególnymi parametrami opisującymi przepływ.
EN
The paper describes results of unsteady flow tests conducted in the pipe supplied with the pulsating flow. Tests were made for two different flow receivers (elements closing the pipe); i.e.: a large volume tank and a turbocharger turbine, with two different pipe lengths. Advanced measurement techniques were applied to record temporal variations of the flow parameters: piesoresistive transducers for pressure measurements, a Constant Current Thermometer for temperature measurements and a Hot Wire probe for fluid velocity measurements. As the effect of these tests, it was found, that the character of flow phenomena for two tested systems (with a tank and a turbine) was different. The correlations between flow parameters were also shown.
PL
Jedna z największych trudności występujących w pomiarach niestacjonarnych temperatur polega na tym, iż własności dynamiczne dostępnych czujników są z reguły niewystarczające do przenoszenia sygnałów w odpowiednio szerokim paśmie częstości. Zwykle stosuje się w takich przypadkach procedurę identyfikacji własności dynamicznych przetwornika a następnie metodę ich korekcji. W artykule zaproponowano inną koncepcję wyznaczania temperatury gazu w oparciu o pomiary wykonywane trójspoinową sondą temperatury. Istotą proponowanej metody jest zmierzenie tego samego sygnału temperatury za pomocą trzech spoin (każda o innej inercyjności), a następnie odtworzenie rzeczywistego sygnału temperatury na podstawie zmierzonych amplitud i faz. W artykule przedstawiono przykładowe wyniki pomiarów oraz przeprowadzono dyskusję na temat ich poprawności oraz aplikacyjności metody.
EN
One of the biggest problems in unsteady temperature measurements is, that dynamic properties of the sensor are insufficient to capture rapid variations of the signal in satisfactory range of frequencies. Usually the procedure of identification and correction of the sensor dynamics is necessary. In the paper, another conception was presented. The idea is based on three junction temperature probes of different thermal inertia. An analysis of amplitudes and phase shifts of the three signals providing from that probe should make possible the reconstruction of the real temperature signal. Some experimental results were presented together with the discussion concerning the applications and the limits of the method.
first rewind previous Strona / 2 next fast forward last