A contactless laser hygrometer based on light absorption by H2O molecules at 1392.5 nm is described. However, measurement results can be affected by optical noise when applied to an atmospheric tunnel or glass cuvette. The noises (occurring in the form of periodic fringes in the recorded spectrum) come from unexpected interference of the light beams reflected from surfaces of the windows or other optical elements. The method of their suppression is described in this article. It is based on wavelength modulation and signal averaging over the fringes period. Also, an experiment confirming the usefulness of this method is described here.
We describe construction and investigation results of optical trace gas sensor working in the 3.334-3.337 μm spectral range. Laser spectroscopy was performed with a multipass cell. A setup was elaborated for detection of ethane at the 3.3368035 μm absorption line. Analysis of the gas spectra and the experiment have shown that, beside C2H6, the sensor is suitable for simultaneous detection of methane, formaldehyde and water vapor. Due to nonlinearity of the laser power characteristic we decided to detect the fourth harmonic of signal. Additional laser wavelength modulation was applied for optical interference suppression. In result, the precision of ethane detection of approximately 80 ppt has been achieved for the averaging time of 20 seconds. Long-term stability as well as the measurement linearity have also been positively tested. The system is suitable for detecting potential biomarkers directly in human breath.