Influence of the transducer bandwidth on compressed ultrasonic echoes

Autorzy
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
The aim of this work is to explore and analyze the influence of the transducer bandwidth on the compressed echoes resulted from the Golay complementary codes transmission. For that reason, a computer simulation and experimental verification were performed that reflected the influence of the transducer bandwidth on the distortion of a signal. This study helps to elucidate why the echoes ringing is present for narrow bandwidth transducers. As known, the shape and symmetry of the pulse waveform and its interaction with the transducer bandwidth and its tuning circuitry have a profound effect on the pulse echo performance achievable from a medical scanning probe. The computer simulation was performed using the Matlab ® software for different fractional transducer bandwidths - from wideband transducers of 100% (ideal case), 90%, 75% to narrowband ones of 50% and 25%. The 16-bits Golay complementary sequences at nominal frequency of 1 MHz were used to illustrate the transducer bandwidth inuence on the resulted signal. It was shown that the decreasing of the transducer bandwidth results in a considerable drop of the amplitude of the compressed echoes from 20.1 V for the 90% fractional bandwidth down to: 17.1 V, 12.5 V and 6.6 V for 75%, 50% and 25% bandwidths, respectively. The widths of the compressed echoes were widening at the same time from 708 ns up to 2.38 s reducing the axial resolution from about 1 mm to over 3.6 mm. In the experiments, two transducers with different fractional bandwidths of 70% and 35% and nominal frequencies of 4.8 MHz and 6 MHz, respectively, were used.
Słowa kluczowe
Rocznik
Strony
903--915
Opis fizyczny
Bibliogr. 10 poz., rys.
Twórcy
autor
  • Institute of Fundamental Technological Research Polish Academy of Sciences Świętokrzyska 21, 00-049 Warszawa, Poland
autor
  • Institute of Fundamental Technological Research Polish Academy of Sciences Świętokrzyska 21, 00-049 Warszawa, Poland
Bibliografia
  • [1] MCKEIGHEN R., Influence of pulse drive shape and tuning on the broadband response of a transducer, Proc. IEEE Ultrasonics Symposium, 2, 1637-1642 (1997).
  • [2] NOWICKI A., LITNIEWSKI J., SECOMSKI W., TROTS I., LEWANDOWSKI M., TYMKIEWICZ R., Coded ultrasonography, Annual Report, Polish Academy of Sciences, 56-58 (2005).
  • [3] NOWICKI A., SECOMSKI W., TROTS I., LITNIEWSKI J., Extending penetration depth using coded ultrasonography, Bulletin of the Polish Academy of Sciences, Technical Sciences, 52, 3, 215-220 (2004).
  • [4] NOWICKI A., TROTS I., LITNIEWSKI J., SECOMSKI W., Golay's codes sequences in ultrasonography, Hydroacoustics, 7, 173-182 (2004).
  • [5] RITTER T., GENG X., SHUNG K., LOPATH P. D., PARK S., SHROUT T. R., Single crystal PZN/PTpolymer composites for ultrasound transducer applications, IEEE Trans. Ultrason. Ferroelect. Freq. Contr., 47, 4, 792-800 (2000).
  • [6] SILVERMAN R. H., VINARSKY E., WOODS S. M., LIZZI F. L., COLEMAN D. F., The effect of transducer bandwidth on ultrasonic image characteristics, Retina, 15, 37-42 (1995).
  • [7] SINGH A., SCHARER, J., BOOSKE J., WOHLBIER J., Second- and third-order signal predistortion for nonlinear distortion suppression in a TWT, IEEE Trans. El. Dev., 52, 5, 709-717 (2005).
  • [8] THAM M. T., Plotting frequency responses, Depth. of Chemical and Process Engineering, University of Newcastle upon Tyne, April 1999.
  • [9] TROTS I., Coded excitation and echoes compression in ultrasonography, Ph.D. Thesis, Ultrasound Department, Institute of Fundamental Technological Research, February 2005.
  • [10] TROTS I., NOWICKI A., SECOMSKI W., LITNIEWSKI J., Golay sequences – side-lobe - canceling codes for ultrasonography, Archives of Acoustics, 29, 1, 87-97 (2004).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-e4e06901-3571-4df9-a44c-441f55d81f00
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.