Identyfikatory
Warianty tytułu
Języki publikacji
Abstrakty
A metrological verification of a high precision digital multimeter was made by the laboratory of calibration of programmable electrical multifunction instruments of the National Institute of Metrological Research (INRIM) in order to verify its accuracy and stability. The instrument had been tested for a period of six months for five low-frequency electrical quantities (DC and AC Voltage and Current and DC Resistance). Its stability and precision were compared with the accuracy specifications of the manufacturer. As a new approach, a performance index of the DMM was introduced and evaluated for each examined measurement point. The DMM showed a satisfactory agreement with its specifications to be considered at the level of other top-class DMMs and even better in some measurements points.
Czasopismo
Rocznik
Tom
Strony
283--296
Opis fizyczny
Bibliogr. 6 poz., rys., tab., wykr.
Twórcy
autor
- National Institute of Metrological Research (INRIM), Strada delle Cacce, 91–10135 Turin, Italy
autor
- National Institute of Metrological Research (INRIM), Strada delle Cacce, 91–10135 Turin, Italy
Bibliografia
- [1] Cassiago, C., La Paglia, G., Pogliano, U. (2000). Stability Evaluation of High-Precision Multifunction Instruments for Traceability Transfer. IEEE Trans. Meas., 49(6) 1206-1210.
- [2] Galliana, F., Capra, P.P., Cerri, R., Lanzillotti, M. (2018). Automated precision DC voltage fixed ratios divider. Measurement, 122, 291-296.
- [3] Galliana, F., Capra, P.P., Cerri, R., Lanzillotti, M. (2018). Automated setup to accurately calibrate electrical dc voltage generators. Measurement, 123, 1291-134.
- [4] Galliana, F., Cerri, R., Roncaglione Tet, L. (2017). High perfomance selectable value transportable high dc Voltage standard. Measurement, 102, 131-137.
- [5] Bich, W., Pennecchi, F. (2004). On the in-use uncertainty of an instrument. P. Ciarlini, M.G., Cox,E., Filipe, F., Pavese, D., Richter (eds). Advanced Mathematical & Computational Tools in Metrology, World Scientific, Singapore, 59-169.
- [6] Galliana, F., Lanzillotti, M. (2017). Accurate comparison between INRIM and a secondary calibration laboratory using a top-class multifunction electrical calibrator. Measurement, 103, 353-360.
Uwagi
PL
Opracowanie rekordu w ramach umowy 509/P-DUN/2018 ze środków MNiSW przeznaczonych na działalność upowszechniającą naukę (2019).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-e1f4ee69-fd49-481e-85e3-7d4edadd3d64