Kalibracja przenośnego spektrometru fluorescencji rentgenowskiej (XRF) na granitach zawierających Sn-W ze złoża Cínovec (Erzgebirge/Krušné Hory Mts., Republika Czeska)
Języki publikacji
Abstrakty
This study evaluates the suitability of portable X-Ray fluorescence spectrometry (PXRF) in the environment of granites and greisens with Sn-W and secondary mineralization of other important elements (Rb,Y, La, Ce, Nb, Ta and Tl). International certificated reference materials (CRM), historical data from the sixties based on the borehole CS-1, current analyses in accredited laboratory and data measured by a portable X-ray fluorescence spectrometer (PXRF) with factory settings were evaluated in this study. Powder CRMs and samples from borehole CS-1 were analysed by PXRF. The results of PXRF analysis were then tested for reliability, credibility, accuracy and precision. The accuracy was calculated as a relative percent difference (RPD), regression equations and correlation coefficients. A detailed statistical evaluation of the results proves, that PXRF can be a very useful method for a primary determination of main, secondary and trace elements. Comparison of the conventional methods and the PXRF shows a good correlation of different analytical methods and a good possibility of using the PXRF method for a future selection of samples for subsequent more demanding and expensive conventional and special analyses. A calibration of the PXRF spectrometer to the lithological environment of Sn-W ore-bearing granites and greisens was made by the statistical comparison of the methods.
Niniejsze badania sprawdzają przydatność przenośnego spektrometru fluorescencji rentgenowskiej (ang. skrót PXRF) w warunkach występowania granitów i grejzenów z Sn-W oraz wtórnej mineralizacji innych pierwiastków (Rb, Y, La, Ce, Nb, Ta i Tl). W badaniach uwzględniono międzynarodowe certyfikowane materiały odniesienia (ang. skrót CRM), dane historyczne z lat 60 oparte na odwiertach CS-1, bieżące analizy w akredytowanych laboratoriach i dane uzyskane z przenośnego spektrometru fluorescencji rentgenowskiej (PXRF) z ustawieniami fabrycznymi. CRM proszkowe oraz próbki pobrane z odwiertu CS-1 przeanalizowano w PXRF. Wyniki analizy PXRF zostały następnie poddane testom na niezawodność, wiarygodność, dokładność i precyzję. Dokładność została policzona jako względna różnica procentowa (RPD), równania regresyjne i współczynniki korelacji. Szczegółowa ocena statystyczna wyników dowodzi, że PXRF może być pomocną metodą we wstępnym określaniu zawartości pierwiastków głównych, wtórnych i śladowych. Porównanie metod konwencjonalnych z metodą PXRF pokazało pozytywną korelację różnych metod analitycznych i możliwość stosowania metody PXRF do przyszłych doborów próbek w celu dalszych trudniejszych i droższych analiz konwencjonalnych i specjalistycznych. Kalibracja spektrometru PXRF do warunków występowania litu w granitach i grejzenach zawierających rudę Sn-W została sporządzona została na podstawie statystycznego porównania wymienionych metod.
Czasopismo
Rocznik
Tom
Strony
67--72
Opis fizyczny
Bibliogr. 16 poz., tab., wykr.
Twórcy
autor
- Czech Geological Survey, Geologická 6, 152 00 Praha 5, Czech Republic
autor
- Czech Geological Survey, Geologická 6, 152 00 Praha 5, Czech Republic
autor
- Czech Geological Survey, Geologická 6, 152 00 Praha 5, Czech Republic
autor
- Institute of Geology AS CR, Rozvojová 269, 165 00 Praha 6, Czech Republic
Bibliografia
- 1. GAZLEY, M.F.; VRY, J.K.; DU PLESSIS, E.; HANDLER, M.R. 2011. "Application of hand-held X-ray fluorescence analyses to metabasalt stratigraphy, Plutonic Gold Mine, Western Australia." Journal of Geochemical Exploration 110: 74–80.
- 2. GAZLEY, M.F.; DUCLAUX, G.; FISHER, L.A.; DE BEER, S.; SMITH, P.; TAYLOR, M.; SWANSON, R.; HOUGH, R.M.; CLEVERLEY, J.S. 2012. "3D visualisation of portable X-ray fluorescence data to improve geological understanding and predict metallurgical performance at Plutonic Gold Mine, Western Australia." Applied Earth Science Trans. 120: 88–96.
- 3. FISHER, L.A.; GAZLEY, M.F.; BAENSCH, A.; BARNES, S.J.; CLEVERLEY, J.; DUCLAUX, G. 2014. "Resolution of geochemical and lithostratigraphic complexity: A workflow for application of portable X-ray fluorescence to mineral exploration." Geochemistry: Exploration, Environment, Analysis 14: 149-159.
- 4. LE VAILLANT, M.; BARNES, S.J.; FISHER, L.; FIORENTINI, M.L.; CARUSO, S. 2014. "Use and calibration of portable X-ray fluorescence analysers: Application to lithogeochemical exploration for komatiite-hosted nickel sulphide deposits." Geochemistry: Exploration, Environment Analysis 14: 199–209.
- 5. ŠTEMPROK, M.; ŠULCEK, Z. 1969. "Geochemical profile through an ore-bearing lithium granite." Economic Geology 64: 392–404.
- 6. RUB, A.K.; ŠTEMPROK, M.; RUB, M.G. 1998. "Tantalum mineralization in the apical part of the Cinovec (Zinnwald) granite stock." Mineralogy and Petrology 63: 199–222.
- 7. JOHAN, Z.; JOHAN, V. 2005. "Accessory minerals of the Cı´novec (Zinnwald) granite cupola, Czech Republic: indicators of petrogenetic evolution." Mineralogy and Petrology 83: 113–150.
- 8. 1994. "Govindaraju. Compilation of working values and sample description for 383 geostandards." Geostandard Newsletter 18: 1–158.
- 9. U.S. EPA. XRF technologies for measuring trace elements in soil and sediment. Niton XLt 700 Series XRF Analyzer. Innovative technology verification report EPA/540/R-06/004. Wasington, 2006.
- 10. BERNICK, MB.; GETTY, D; PRINCE, G.; SPRENGER, M. 1995. "Statistical evaluation of field-portable X-ray fluorescence soil preparation methods." Journal of Hazardous Materials 43: 111–116.
- 11. HEWITT, A.D. 1995. "Rapid screening of metals using portable high resolution X-ray fluorescence spectrometers." Available online: <http://www.crrel.usace.army.mil/techpub/CRREL_Reports/reports/SR95_14.pdf>.
- 12. KALNICKY, JD.; SINGHVI, R. 2001. "Field portable XRF analysis of environmental samples." Journal of Hazardous Materials 83: 93–122.
- 13. MÄKINEN, E.; KORHONEN, M.; VISKARI, EL.; HAAPAMÄKI, S.; JÄRVINEN, M.; LU, L. 2005. Comparison of XRF and FAAS methods in analysing CCA contaminated soils. Water, Air, and Soil Pollution 171: 95–110.
- 14. KILBRIDE, C.; POOLE, J.; HUTCHINGS, T.R. 2006. "A comparison of Cu, Pb, As, Cd, Zn, Fe, Ni and Mn determined by acid extraction/ICP–OES and ex situ field portable X-ray fluorescence analyses." Environmental Pollution 143: 16–23.
- 15. CERNUSCHI, F.; FORD, M.T.; BOSCHMANN, D.; DILLES, J.H.; CONREY, R.M. 2013. "Test of a user defined calibration in a portable XRF for lithogeochemistry applications." in: E. Jonsson et al (eds.) "Mineral deposit research for high-tech world. Proceedings of the 12th Biennial SGA Meeting" Geological Survey of Sweden 1: 180–183.
- 16. SIMANDL, G.J.; STONE, R.S.; PARADIS, S.; FAJBER, R.; REID, H.M.; GRATTAN, K. 2014. "An assessment of a handheld X-ray fluorescence instrument for use in exploration and development with an emphasis on REEs and related specialty metals." Mineralium Deposita 49: 999–1012.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-d55940f6-b5a5-4bcb-85a4-a97fa0ff1bb3