PL EN


Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników
Tytuł artykułu

An experimentally validated model for the ilizarov fixator considering the loss of wire’s pretension

Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
The purpose of this study is to assess the biomechanical performance of the Ilizarov fixator and to attempt optimization of its configuration. Attention is focused to the quantification of the wires’ pretension loss. The study is implemented according to a combined experimental and numerical scheme: The data of a long series of experiments are used to calibrate and validate a numerical model. The model is then used for parametric analysis of the factors which, according to clinical experience, influence the efficiency of the device. It was concluded that the most critical factors are the pretension of the wires, the angle between them and the rings’ diameter; the role of the remaining factors is of rather minor importance.
Rocznik
Strony
621--627
Opis fizyczny
Bibliogr. 4 poz., rys., wykr.
Twórcy
  • National Technical University of Athens Department of Mechanics Laboratory of Testing and Materials Theocaris Building, Zografou Campus, 15773, Athens, Greece
  • National Technical University of Athens Department of Mechanics Laboratory of Testing and Materials Theocaris Building, Zografou Campus, 15773, Athens, Greece
Bibliografia
  • 1. Fleming B., Paley D., Kristiansen T., Pope M., A biomechanical analysis of the Ilizarov external fixator, Clinical Orthopaedics and Related Research, 241: 95–105, 1989.
  • 2. Mitousoudis A.S., A study of the Ilizarov external osteosynthesis in the direction of documenting optimum assemblies for various clinical scenarios (url: http://dspace.lib.ntua.gr/handle/123456789/3761), PhD Thesis, National Technical University of Athens, Department of Mechanics, Supervisor: S.K. Kourkoulis, 2011.
  • 3. Ryan S., Ehrhart N., Zuehlsdorff K., James S., Comparison of alternate and simultaneous tensioning of wires in a single-ring fixator construct, Veterinary Surgery, 38(1): 96–103, 2009, doi: 10.1111/j.1532-950X.2008.00469.x.
  • 4. Kourkoulis S.K., Mitousoudis A.S., A parametric study of the Ilizarov fixator with an experimentally validated model considering the loss of wire’s pretension, to be submitted, 2016.
Uwagi
Opracowanie ze środków MNiSW w ramach umowy 812/P-DUN/2016 na działalność upowszechniającą naukę (zadania 2017).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-d2aaf0e4-4ed8-4e7b-8c95-e034a6840ed1
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.