Evaluation of the reliability of a new non-invasive method for assessing the functionality and mobility of the spine

Treść / Zawartość
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
For the evaluation of the functionality and mobility of the spine, several methods have been developed. The purpose of this study was to estimate the test-retest reliability of the Spinal Mouse, a new, non-invasive, computer-assisted wireless telemetry device for the assessment of the curvatures, the mobility and the functionality of the spine. Materials and methods: the test-retest reliability was evaluated in 50 adults with back or low back pain. Twenty four parameters were studied in the sagittal and frontal plane. For the characterization of the precision, the intraclass correlation coefficient and the standard error of measurement were used. Results: in the sagittal plane, 22 of the 24 parameters showed high and good reliability, while only two fair and poor. In the frontal plane, 17 parameters showed high and good reliability, five fair and two poor. Discussion: the Spinal Mouse showed excellent test-retest reliability in the sagittal plane, while a slightly inferior performance in the frontal plane, for the evaluation of curvatures, deformation and mobility of the spine.
Rocznik
Strony
117--124
Opis fizyczny
Bibliogr. 37 poz., rys., tab.
Twórcy
autor
  • University of Crete – Faculty of Medicine, Department of Orthopaedics and Traumatology, University Hospital of Heraklion
  • Alexander S. Onassis Public Benefit Foundation
  • University of Crete – Faculty of Medicine, Department of Orthopaedics and Traumatology, University Hospital of Heraklion
autor
  • Department of Anaesthesiology, University Hospital of Heraklion, Crete, Greece
autor
  • University of Crete – Faculty of Medicine, Department of Orthopaedics and Traumatology, University Hospital of Heraklion
autor
  • University of Crete – Faculty of Medicine, Department of Orthopaedics and Traumatology, University Hospital of Heraklion
Bibliografia
  • [1] SALISBURY P.J., PORTER R.W., Measurement of lumbar sagittal mobility. A comparison of methods, Spine, 1987, Vol. 12, 190–193.
  • [2] YOUDAS J.W., SUMAN V.J., GARRETT T.R., Reliability of measurements of lumbar spine sagittal mobility obtained with the flexible curve, J. Orthop. Sports Phys. Ther., 1995, Vol. 21, 13–20.
  • [3] HARD F.D., STRICKLAND D., CLIFFE P., Measurement of spinal mobility, Ann. Rheum. Dis, 1974, Vol. 33, 136–139.
  • [4] KOROVESSIS P., KOUREAS G., PAPAZISIS Z., Correlation between backpack weight and way of carrying, sagittal and frontal spinal curvatures, athletic activity, and dorsal and low back in schoolchildren and adolescents, J. Spinal Disord. Tech., 2004, Vol. 17(1), 33–40.
  • [5] GILL K., KRAG M.H., JOHNSON G.B., HAUGH L.D., POPE M.H., Repeatability of four clinical methods for assessment of lumbar spinal motion, Spine (Philadelphia Pa 1976), 1988, Vol. 13(1), 50–53.
  • [6] NORTON B.J., KELLY H., ZOU D., Comparisons among non-invasive methods for measuring lumbar curvature in standing, J. Orthop. Sports Phys. Ther., 2002, Vol. 32(8), 405–414.
  • [7] TILITSON K.M., BURTON A.K., Non-invasive measurement of lumbar sagittal mobility. An assessment of the flexicurve technique, Spine, 1991, Vol. 16, 29–34.
  • [8] VIITANEN J.V., KOKKO M.L., HEIKKILA S., KAUTIAINEN H., Neck mobility assessment in ankylosing spondylitis: A clinical study on nine measurements including new tape method for cervical rotation and lateral flexion, Br. J. Rheumatol., 1988, Vol. 37, 377–381.
  • [9] HALEY S.M., TADA W.L., CARMICHAEL E.M., Spinal mobility in young children, Physical Therapy, 1986, Vol. 66(11), 1697–1705.
  • [10] MANNION A.F., KNECHT K., BALABAN G., DVORAL J., GROB D., A new skin-surface device for measuring the curvature and global and segmental ranges of motion of the spine: reliability of measurements and comparison with data reviewed from literature, Eur. Spine J., 2004, Vol. 13, 122–136.
  • [11] CHO K.J., RAH U.W., Study of the lumbar curvature with various factors of pelvic inclination, Yonsei Medical Journal, 1995, Vol. 36(2), 153–160.
  • [12] POST R.B., LEFERINK V.J.M., Spinal mobility: sagittal range of motion measured with the Spinal Mouse, a new noninvasive device, Arch. Orthop. Trauma Surg., 2004, Vol. 124, 187–192.
  • [13] NAKIPOGLU G.F., KARAGÖZ A., ÖZGIRGIN N., The biomechanics of lumbosacral region in acute and chronic low back pain patients, Pain Physician, 2008, Vol. 11(4), 505–511.
  • [14] GORI T., MÜNZEL T., Biological effects of low-dose radiation: of harm and hormesis, European Heart Journal, 2012, Vol. 33, 292–295.
  • [15] DAGAL A., Radiation safety for anesthesiologists, Curr. Opin. Anaesthesiol., 2011, Vol. 24(4), 445–450.
  • [16] BRENNER D.J., Should we be concerned about the rapid increase in CT usage?, Rev. Environ. Health, 2010, Vol. 25(1), 63–68.
  • [17] de GONZALEZ A.B., DARBY S., Risk of cancer from diagnostic X-rays: estimates for the UK and 14 other countries, Lancet, 2004, Vol. 363, 345–351.
  • [18] BRENNER D.J., ELLISTON C.D., Estimated radiation risks potentially associated with full-body CT screening, Radiology, 2004, Vol. 233, 735–738.
  • [19] GSTOETTNER M., SEKYRA K., WALOCHNIK N., WINTER P., WACHTER R., BACH C.M., Inter- and intraobserver reliability assessment of the Cobb angle: manual versus digital measurement tools, Eur. Spine J., 2007, Vol. 16(10), 1587–1592.
  • [20] CHAKRAVERTY R., PYNSENT P., ISAACS K., Which spinal levels are identified by palpation of iliac crest and the posterior superior iliac spines?, J. Anatom., 2007, Vol. 210, 232–236.
  • [21] GRAVES J.E., POLLOCK M.L., CARPENTER D.M., LEGGETT S.H., JONES A., MACMILLAN M., FULTON M., Quantitative assessment of full range-of-motion isometric lumbar extension strength, Spine (Philadelphia Pa 1976), 1990, Vol. 15(4), 289–294.
  • [22] SEVERINSSON Y., ELISSON L., BUNKETORP O., Reliability of measuring the cervical sagittal translation mobility with a simple method in a clinical setting, Rehabilitation Research and Practice, 2012, (in press) ID: 629104, DOI: 10.1155/2012/629104.
  • [23] PERNEGER T.V., What’s wrong with Bonferroni adjustments?, Br. Med. J., 1998, Vol. 316, 1236–1238.
  • [24] WEIR J., Quantifying test-retest reliability using the Intraclass Correlation Coefficient and the SEM, J. Strength Cond. Res., 2005, Vol. 19(1), 231–240.
  • [25] CHINN S., Repeatability and method comparison, Thorax, 1991, Vol. 46, 454–456.
  • [26] FENTER P.C., BELLEW J.W., PITTS T.A., KAY R.E., Reliability of stabilised commercial dynamometers for measuring hip abduction strength: a pilot study, Br. J. Sports Med., 2003, Vol. 37, 331–334.
  • [27] WILLIAMS J.M., HAQ I., LEE R.Y., Dynamic lumbar curvature measurement in acute and chronic low back pain sufferers, Arch. Phys. Med. Rehabil., 2012, Vol. 93, 2094–2099.
  • [28] VAZ G., ROUSSOULY P., BERTHONNAUD E., DIMNET J., Sagittal morphology and equilibrium of pelvis and spine, Eur. Spine J., 2002, Vol. 11, 80–87.
  • [29] CARLUCCI L., CHIU J.C., CILIFFORD T.J., Spinal mouse for assessment of spinal mobility, The Internet Journal of Minimally Invasive Spinal Technology, 2001, Vol. 1, 30–31.
  • [30] KELLIS E., ADAMOU G., TZILIOS G., EMMANOUILIDOU M., Reliability of spinal range of motion in healthy boys using a skin-surface device, Journal of Manipulative and Physiological Therapeutics, 2008, Vol. 31(8), 570–576.
  • [31] KELLER S., MANNION F.A., GROB D., Reliability of a new measuring device “Spinal Mouse” in recording the sagittal profile of the back, Eur. Spine J., 2000, Vol. 9, 303.
  • [32] BAUTMANS I., VAN ARKEN J., VAN MACKELENBERG M., METS T., Rehabilitation using manual mobilization for thoracic kyphosis in elderly postmenopausal patients with osteoporosis, J. Rehabil. Med., 2010, Vol. 42, 129–135.
  • [33] BRUTON A., CONWAY J.H., HOLGATE S.T., Reliability: What is it, and how it is measured?, Physiotherapy, 2000, Vol. 86, 94–99.
  • [34] FLEISS J.L., The design and analysis of clinical experiments, New York, NY, Wiley, 1986.
  • [35] SLEIVERT G.G., WENGER H.A., Reliability of measuring isometric and isokinetic peak torque, rate of torque development, integrated electromyography, and tibial nerve conduction velocity, Arch. Phys. Med. Rehabil., 1994, Vol. 75, 1315–1321.
  • [36] CHEN S.P., SAMO D.G., CHEN E.H., CRAMPTON A.R., CONRAD K.M., EGAN L., MITTON J, Reliability of tree lumbar sagittal motion measurement methods: surface inclinometers, J. Occup. Environ. Med., 1997, Vol. 39, 217–223.
  • [37] DOPF C.A., MANDEL S.S., GEIGER D.F., MAYER P.J., Analysis of spine motion variability using a computerized goniometer compered to physical examination. A prospective clinical study, Spine, 1994, Vol. 19, 586–595.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-c4e2d3c3-bd09-43f1-81a3-6f5cd4f96555
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.