Effects of the pre-irradiation storage procedure on the dose response of a Fricke xylenol orange gel dosimeter

Treść / Zawartość
Identyfikatory
Warianty tytułu
Konferencja
International Conference on Development and Applications of Nuclear Technologies NUTECH 2014 (21-24.09.2014, Warsaw, Poland)
Języki publikacji
EN
Abstrakty
EN
The Fricke xylenol orange (FX) gel system is a chemical dosimeter characterized by good sensitivity, linear dose response, tissue equivalence, no toxicity, easy preparation, reproducibility and low cost. Thanks to the presence of the gelatinous matrix, the system is particularly suitable to perform reliable 3D mapping of the absorbed dose spatial distribution via magnetic resonance imaging (MRI) or optical techniques. The aim of this work is to study in a systematic way the influence of the pre-irradiation storage procedure upon sensitivity, dose response stability and lifetime of use of a FX gel system made with gelatin from porcine skin subjected to homogeneous irradiation. For this purpose, different pre-irradiation storage procedures, in terms of temperature and duration of each storage step, were investigated. In order to evaluate the dose response stability, the optical analyses of the samples were performed up to 6 hours after irradiation. Moreover, the samples were irradiated at time intervals of 24 hours for up to 7 days after preparation in order to evaluate the system lifetime of use. Regardless of their thermal and temporal life, the samples show linear dose responses in the investigated dose range (3–24 Gy) and an increase of sensitivity with the time elapsed between preparation and irradiation. Among the three pre-irradiation storage procedures considered here, a procedure that provides the best dose response stability and lifetime of use was identifi ed and recommended for further use. The analyzed dosimetric system possesses good properties that make it promising for medical application, particularly concerning the evaluation of pre-treatment plan quality assurance within the conformational external beam radiotherapy.
Czasopismo
Rocznik
Strony
603--607
Opis fizyczny
Bibliogr. 14 poz., rys.
Twórcy
autor
  • Department of Energy, Nuclear Engineering Division-CeSNEF, Politecnico di Milano, Piazza L. da Vinci 32, Milano, I-20133, Italy, Tel.: 02239 9635
autor
  • Department of Energy, Nuclear Engineering Division-CeSNEF, Politecnico di Milano, Piazza L. da Vinci 32, Milano, I-20133, Italy, Tel.: 02239 9635
autor
  • Medical Physics Unit, Fondazione IRCCS, Istituto Nazionale dei Tumori, via Venezian, Milano, I-20113, Italy
autor
  • Medical Physics Unit, Fondazione IRCCS, Istituto Nazionale dei Tumori, via Venezian, Milano, I-20113, Italy
autor
  • Department of Physics, Università degli Studi di Milano and INFN, via Celoria 16, Milano, I-20133 Italy
autor
  • Department of Energy, Nuclear Engineering Division-CeSNEF, Politecnico di Milano, Piazza L. da Vinci 32, Milano, I-20133, Italy, Tel.: 02239 9635
Bibliografia
  • 1. Attix, F. H. (2004). Introduction to radiological physics and radiation dosimetry. Germany: Wiley--VCH Verlag.
  • 2. Appleby, A., & Leghrouz, A. (1991). Imaging of radiation dose by visible color development in ferrousagarose Xylenol orange gels. Med. Phys., 18, 309–312.DOI: 10.1118/1.596676.
  • 3. Lepage, M., & Jordan, K. (2010). 3D dosimetry fundamentals: gels and plastics. J. Phys.-Conf. Ser., 250,253–261. DOI: 10.1088/1742-6596/250/1/012055.
  • 4. Schreiner, L. J. (2004). Review of Fricke gel dosimeters. J. Phys.-Conf. Ser., 3, 9–21. DOI: 10.1088/1742-6596/3/1/003.
  • 5. Keall, P., & Baldock, C. (1999). A theoretical study of the radiological properties and water equivalence of Fricke and polymer gels used for radiation dosimetry.Austral. Phys. Eng. Sci. Med., 22(3), 85–91.
  • 6. Davies, J. B., & Baldock, C. (2008). Sensitivity and stability of the Fricke-gelatin-xylenol orange gel dosimetery. Radiat. Phys. Chem., 77(6), 690–696.DOI: 10.1016/j.radphyschem.2008.01.007.
  • 7. Bero, M. A. (2007). Dosimetric properties of a radiochromic gel detector for diagnostic X-rays. Nucl. Instrum. Methods Phys. Res. Sect. A-Accel. Spectrom.Dect. Assoc. Equip., 39(7), 186–189. DOI: 10.1016/j.nima.2007.05.080.
  • 8. Gambarini, G., Gomarasca, G., Marchesini, R.,Pecci, A., Pirola, L., & Tomatis, S. (1999). Three-dimensional determination of absorbed dose by spectrophotometric analysis of ferrous-sulphate agarose gel. Nucl. Instrum. Methods Phys. Res. Sect. A-Accel. Spectrom. Dect. Assoc. Equip., 422(1/3), 643–648.DOI: 10.1016/S0168-9002(98)00975-9.
  • 9. Baldock, C., Harris, P. J., Piercy, A. R., & Healy, B. (2001). Experimental determination of the diffusion coefficient in two-dimensions in ferrous sulphate gels using the fi nite element method. Austral. Phys. Eng. Sci. Med., 24(1), 19–30. DOI: 10.1007/BF03178282.
  • 10. Gore, J. C., & Kang, Y. S. (1984). Measurement of radiation dose distributions by nuclear magnetic resonance (NMR) imaging. Phys. Med. Biol., 29(10), 1189–1197. DOI: 10.1088/0031-9155/29/10/002.
  • 11. De Deene, Y. (2004). Fundamentals of MRI measurements for gel dosimetry. J. Phys.-Conf. Ser., 3, 87–114.DOI: 10.1088/1742-6596/3/1/009.
  • 12. Jordan, K. J., & Battista, J. (2004). Dose response of ferrous-xylenol orange gels: the effects of gel substrate, gelation time and dose fractionation. J. Phys.-Conf. Ser., 3, 232–235. DOI: 10.1088/1742-6596/3/1/035.
  • 13. Jordan, K. J., MacDonald, D. J., Pajak, T., & Battista, J. J. (2001). Refrigeration temperature gradients generate dose response artifacts in Fricke gelatin dosimeters. In DosGel2001 2nd International Conference on Radiotherapy Gel Dosimetry (pp. 95–98).Brisbane, Australia.
  • 14. Davies, J. B., & Baldock, C. (2010). Temperature dependence on the dose response of the Fricke-gelatin-xylenol orange gel dosimeter. Radiat. Phys. Chem., 79(5), 660–662. DOI: 10.1016/j.radphyschem. 2009.11.014.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-b4730663-2d4d-4bb0-a487-d9b5f6320ae4
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.