PL EN


Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników
Powiadomienia systemowe
  • Sesja wygasła!
  • Sesja wygasła!
Tytuł artykułu

Calibration of Gafchromic XR-RV3 film under interventional radiology conditions

Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
Introduction: The purpose of the study was the calibration of Gafchromic films in clinical interventional radiology conditions and the assessment of the influence of dose range, the shape of the fitting curve, and its practical application. The aim of the work was to show how practically perform calibration in a wide range of doses. Material and methods: Gafchromic XR–RV3 films were included in the study. The calibration was performed for A and B film series separately. Doses from the range of 0 – 8 Gy were used. Film dosimeters were read out in reflective mode with a commercial flatbed scanner. Results: Among various degrees of a polynomial function, the best fit, which fulfilled the chosen criterion of 95% agreement between measured and reconstructed doses and simple equation criterion, was observed for third-degree polynomial. The fitting curve where the dose is the function of optical density (logMPV) was demonstrated to be more precise than the fitting curve based on MPV only. To minimize the difference between dose absorbed by the film and dose reconstructed from the fitting curve below 5% it is necessary to divide the calibration range of 0 – 8 Gy into two subranges for use in interventional radiology. This difference was set at a maximum level of 3.8% and 1.9% for the lowand high-dose range, respectively. Each series of films may have a slightly different calibration curve, especially for the low dose range. A deviation of up to 36% between two batches of Gafchromic film was observed. Conclusions: For the third-degree polynomial fitting function (one of the recommended in the literature) calibration should be done into low and high dose ranges and for each batch separately. A systematic error higher than 20% could be introduced when the fitting curve from one film batch is applied to the other film batch.
Rocznik
Strony
165--173
Opis fizyczny
Bibliogr. 25 poz., rys., tab.
Twórcy
  • Invasive Cardiology and Electrocardiology Department, 3rd Department of Cardiology, Medical University of Silesia, Katowice, Poland
  • Prof. K. Gibiński Memorial University Clinical Centre, Medical University of Silesia, Poland
  • Department of Electrocardiology and Heart Failure, Medical University of Silesia, Poland
  • 3rd Department of Cardiology, School of Medicine in Katowice, Medical University of Silesia, Poland
  • Invasive Cardiology and Electrocardiology Department, 3rd Department of Cardiology, Medical University of Silesia, Katowice, Poland
Bibliografia
  • 1. Farah J, Trianni A, Carinou E, et al. Measurement of maximum skin dose in interventional radiology and cardiology and challenges in the set-up of European alert thresholds. Radiat Prot Dosimetry. 2015;164(1-2):138-142. https://doi.org/10.1093/rpd/ncu314
  • 2. Greffier J, Van Ngoc Ty C, Agelou M, et al. Impact of the calibration conditions of XR-RV3 films on peak skin dose measurements in interventional radiology. Radiat Prot Dosimetry. 2017;174(2):207-215. https://doi.org/10.1093/rpd/ncw116
  • 3. McCabe BP, Speidel MA, Pike TL, Van Lysel MS. Calibration of GafChromic XR-RV3 radiochromic film for skin dose measurement using standardized x-ray spectra and a commercial flatbed scanner. Med Phys. 2011;38(4):1919-1930. https://doi.org/10.1118/1.3560422
  • 4. Farah J, Trianni A, Ciraj-Bjelac O, et al. Characterization of XR-RV3 GafChromic(®) films in standard laboratory and in clinical conditions and means to evaluate uncertainties and reduce errors. Med Phys. 2015;42(7):4211-4226. https://doi.org/10.1118/1.4922132
  • 5. Wen N, Lu S, Kim J, et al. Precise film dosimetry for stereotactic radiosurgery and stereotactic body radiotherapy quality assurance using GafchromicTM EBT3 films. Radiat Oncol. 2016;11(1):1-11. https://doi.org/10.1186/s13014-016-0709-4
  • 6. Girardi A, Anglesio S, Amadore G, Trevisiol E, Redda MG. Implementation of a program for quality assurance on leaf positioning accuracy using Gafchromic ® RTQA2 films. J Med Phys. 2014;39(2):116-120. https://doi.org/10.4103/0971-6203.131287
  • 7. León Marroquin EY, Herrera González JA, Camacho López MA, Villarreal Barajas JE, García-Garduño OA. Evaluation of the uncertainty in an EBT3 film dosimetry system utilizing net optical density. J Appl Clin Med Phys. 2016;17(5):466-481. https://doi.org/10.1120/jacmp.v17i5.6262
  • 8. Paelinck L, De Neve W, De Wagter C. Precautions and strategies in using a commercial flatbed scanner for radiochromic film dosimetry. Phys Med Biol. 2007;52(1):231-242. https://doi.org/10.1088/0031-9155/52/1/015
  • 9. Miller DL, Balter S, Cole PE, et al. Radiation doses in interventional radiology procedures: the RAD-IR study: part I: overall measures of dose. J Vasc Interv Radiol. 2003;14(6):711-727. https://doi.org/10.1097/01.rvi.0000079980.80153.4b
  • 10. Costa KC, Gomez AML, Alonso TC, Mourao AP. Radiochromic film calibration for the RQT9 quality beam. Radiat Phys Chem. 2017;140:370-372. https://doi.org/https://doi.org/10.1016/j.radphyschem.2017.02.032
  • 11. Bin JH, Ji Q, Seidl PA, et al. Absolute calibration of GafChromic film for very high flux laser driven ion beams. Rev Sci Instrum. 2019;90(5):053301. https://doi.org/10.1063/1.5086822
  • 12. Gorny KR, Leitzen SL, Bruesewitz MR, Kofler JM, Hangiandreou NJ, McCollough CH. The calibration of experimental selfdeveloping Gafchromic® HXR film for the measurement of radiation dose in computed tomography. Med Phys. 2005;32(4):1010-1016. https://doi.org/https://doi.org/10.1118/1.1862802
  • 13. Tomic N, Quintero C, Whiting BR, et al. Characterization of calibration curves and energy dependence GafChromicTM XR-QA2 model based radiochromic film dosimetry system. Med Phys. 2014;41(6Part1):62105. https://doi.org/https://doi.org/10.1118/1.4876295
  • 14. Greffier J, Goupil J, Larbi A, et al. Assessment of patient’s peak skin dose during abdominopelvic embolization using radiochromic (Gafchromic) films. Diagn Interv Imaging. 2018;99(5):321-329. https://doi.org/10.1016/j.diii.2017.12.008
  • 15. Balter S. Methods for measuring fluoroscopic skin dose. Pediatr Radiol. 2006;36:136-140. https://doi.org/10.1007/s00247-006-0193-3
  • 16. Chaikh A, Gaudu A, Balosso J. Monitoring methods for skin dose in interventional radiology. Int Journalof Cancer Ther Oncol. 2014;3:3011. https://doi.org/10.14319/ijcto.0301.1
  • 17. Greffier J, Van Ngoc Ty C, Agelou M, et al. Impact of the calibration conditions of XR-RV3 films on peak skin dose measurements in interventional radiology. Radiat Prot Dosimetry. 2017;174(2):207-215. https://doi.org/10.1093/rpd/ncw116
  • 18. Devic S, Tomic N, Lewis D. Reference radiochromic film dosimetry: Review of technical aspects. Phys Medica. 2016;32(4):541-556. https://doi.org/10.1016/j.ejmp.2016.02.008
  • 19. Setilo I, Du Plessis FCP. Dosimetric comparison between XR-RV3 and EBT2 radiochromic film in megavoltage photon beams. Int J Radiat Res. 2016;14(2):149-152. https://doi.org/10.18869/acadpub.ijrr.14.2.149
  • 20. Pasquino M, Cutaia C, Poli M, et al. Patient’s Peak Skin Dose evaluation using Gafchromic films in interventional cardiology procedures and its correlation with other dose indicators. Phys Medica. 2018;53:103-107. https://doi.org/10.1016/j.ejmp.2018.08.017
  • 21. Niroomand-Rad A, Chiu-Tsao S-T, Grams MP, et al. Report of AAPM Task Group 235 Radiochromic Film Dosimetry: An Update to TG-55. Med Phys. 2020;47(12):5986-6025. https://doi.org/https://doi.org/10.1002/mp.14497
  • 22. Jones AK, Ensor JE, Pasciak AS. How accurately can the peak skin dose in fluoroscopy be determined using indirect dose metrics? Med Phys. 2014;41(7):71913. https://doi.org/10.1118/1.4884020
  • 23. Jones AK, Pasciak AS. Calculating the peak skin dose resulting from fluoroscopically guided interventions. Part I: Methods. J Appl Clin Med Phys. 2011;12(4):3670. https://doi.org/10.1120/jacmp.v12i4.3670
  • 24. Thwaites D. Accuracy required and achievable in radiotherapy dosimetry: have modern technology and techniques changed our views? J Phys Conf Ser. 2013;444:12006. https://doi.org/10.1088/1742-6596/444/1/012006
  • 25. Greffier J, Grussenmeyer-Mary N, Larbi A, et al. Experimental evaluation of a radiation dose management system-integrated 3D skin dose map by comparison with XR-RV3 Gafchromic® films. Phys Medica. 2019;66:77-87. https://doi.org/10.1016/j.ejmp.2019.09.234
Uwagi
Opracowanie rekordu ze środków MNiSW, umowa Nr 461252 w ramach programu "Społeczna odpowiedzialność nauki" - moduł: Popularyzacja nauki i promocja sportu (2021).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-b30ee3a8-2a26-4871-ae39-c4d1d5c57dd5
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.