Influence of defibrillation treatment on some electrical parameters and application safety of ECG electrodes

Autorzy
Treść / Zawartość
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
The subject of examination was mainly the determination of safety of available ECG electrodes use in conditions of defibrillation treatment in monitored patients. The electrodes of four European producers used in healthcare have been tested. For reason of periods of electrodes obtaining and their physical condition, the examinations have comprised four selected years and 7 types of electrodes. For testing needs, the electrodes were glued into pairs, in gel-to-gel mode. Testing was based on requirements of globally used American ANSI⁄AAMI EC 12:2000⁄(R)2010 standard for disposable ECG electrodes. For the electrodes pairs tested, determined was the influence of fourfold defibrillation release on polarization potential and their impedance measured with the current of 10 Hz frequency. It has been generally found that in result of fourfold exposition to defibrillating pulse the drop of impedance occurred in all tested, electrode pairs. For each electrodes series, mean value of this drop falls into interval of 10,8% to 57%. It was also pointed out that the ECG electrodes examined, while complying to requirements of the standard, allow the operator to perform the defibrillation procedure safely, including the possibility to fast evaluate the defibrillation result basing on ECG waveform recorded. Such general estimation of safety level of the available European ECG electrodes may be used as some reference in today very diversified market as to provenience end quality of products.
Słowa kluczowe
Rocznik
Tom
Strony
133--139
Opis fizyczny
Bibliogr. 7 poz., rys., tab.
Twórcy
autor
  • Institute of Medical Technology and Equipment ITAM, 118, Roosevelt Str., Zabrze, Poland
autor
Bibliografia
  • [1] ANSI/AAMI EC 12:2000/(R)2010 Disposable ECG electrodes.
  • [2] EN 60601-1:2006 Medical electrical equipment. Part 1: General requirements for basic safety and essential performance.
  • [3] HOI-JUN YOO, VAN HOOF CH. Bio-Medical CMOS ICs (Integrated Circuits and Systems) 2011, Springer, New York
  • [4] KLINGLER DR, SCHOENBERG AA, WORTH NP, EGLESTON CF, BURKART JA. A comparison of gel-to-gel and skin measurements of electrode impedance. Med. Instrum., 1979, Sep-Oct;13(5), pp. 266-268.
  • [5] OLSON WH, SCHMINCKE DR, HENLEY BL. Time and frequency dependence of disposable ECG electrode-skin impedance. Med. Instrum., 1979, Sep-Oct;13(5), pp. 269-272.
  • [6] PATTERSON RP. The electrical characteristics of some commercial ECG electrodes. J Electrocardiol., 1978 Jan;11(1), pp. 23-26
  • [7] SCHOENBERG AA, BOOTH HE, LYON PC. Development of standard test methods for evaluating defibrillation recovery characteristics of disposable ECG electrodes. Med. Instrum., 1979 Sep-Oct;13(5), pp. 259-265.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-PWA4-0026-0015
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.