Implementation of temperature control into a high resolution gamma ray detector for tomography applications

Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
In this paper the development of a new thermal control design for a high resolution gamma ray computed tomography detector is presented. The new design has become necessary due to stringent requirements on count rate accuracy under varying thermal conditions at thermal hydraulic test loops. The components of the existing tomography detector have been analysed with respect to their thermal response and a new arrangement has been devised that is optimized with respect to efficient heat conduction from internal as well as external heat sources to a cooling circuit.
Słowa kluczowe
Rocznik
Tom
Strony
167--172
Opis fizyczny
Bibliogr. 8 poz.
Twórcy
autor
autor
autor
autor
autor
  • Institute of Safety Research, Forschungszentrum Dresden - Rossendorf, Dresden, Germany, a.bieberle@fzd.de
Bibliografia
  • BIEBERLE A., KRONENBERG J., HAMPEL U., (2006), Void fraction distribution measurement in fuel element bundles with a new high-resolution gamma, ATW -International Journal for Nuclear Power Vol. 51 8/9, pp.539 - 540.
  • BIEBERLE A., KRONENBERG J., SCHLEICHER E., HAMPEL U., (2007), Design of a high resolution gamma ray detector module for tomography applications, Nuclear Instruments and Methods in Physic Research A, pp.668 -675.
  • BIEBERLE A., SCHLEICHER E., HAMPEL U., (2007a), Data acquisition system for angle synchronized gamma ray tomography of rapidly rotating objects, Measurement Science and Technology Vol. 18, pp.3384 - 3390.
  • GRAEME J., (1996), Photodiode Amplifiers: Op Amp Solution, McGraw-Hill Publishing Co.
  • HAMPEL U. BIEBERLE A., HOPPE D., KRONENBERG J., SCHLEICHER E., SÜHNEL T., ZIMMERMANN F., ZIPPE C, (2007a), High resolution gamma ray tomography scanner for flow measurement and nondestructive testing applications, Review of Scientific Instruments Vol. 78, 103704.
  • HAMPEL U., HRISTOV H.V., BIEBERLE A., ZIPPE C, (2007), Application of high resolution gamma ray tomography to the measurement of gas hold-up distributions in a stirred chemical reactor, Flow Measurement and Instrumentation, Vol.18, pp. 184 - 190.
  • KAK C, SLANEY M., (1988), Principles of Computerized Tomographic Imaging, IEEE Press, New York.
  • SAINT GOBAIN CRYSTALS, (2007), Datasheet of PreLude 420, Saint Gobain Ceramics & Plastics Inc.