System for fetal magnetocardiography in an unshielded environment

Identyfikatory
Warianty tytułu
PL
System do magnetokardiografii serca płodu w środowisku nieekranowanym
Konferencja
Mechatronics 2004. 5 International Conference ; 23-25.09.2004 ; Warsaw, Poland
Języki publikacji
EN
Abstrakty
EN
This paper presents a developed system designed for fetal magnetocardiography in magnetically unshielded environment. The SQUID based system, equipped with a third order flux transformer, has a capability for recording fetal magnetocardiogram in a hospital setting. The incorporation of the optimally designed third order flux transformer enhances the signal to noise ratio of the magnetometer. An algorithm for optimal configuration of the third order flux transformer is outlines. Spatial characteristics of the third order gradiometer are discussed. Description of the maeasuring system as well as the signal processing procedure for fetal magnetocardiogram recordings are also presented.
PL
Przedstawiono tu system skonstruowany i opracowany w celu pomiaru magnetokardiogramów serca płodu w środowisku nieekranowanym magnetycznie. Oparty o SQUID (Superconducting QUantum Interference Device - Nadprzewodzący Interferometr Kwantowy) i transformator strumienia magnetycznego trzeciego rzędu magnetometr umożliwia pomiar sygnałów magnetycznych pochodzących od aktywności serca płodu w warunkach szpitalnych. Optymalnie zaprojektowany transformator strumienia wydatnie polepsza stosunek sygnału do szumu magnetometru. Opisane zostały więc wytyczne umożliwiające optymalną konfigurację transformatora strumienia magnetycznego trzeciego rzędu. Następnie przedyskutowana została charakterystyka przestrzenna gradiometru trzeciego rzędu. Ponadto przedstawiono system pomiarowy wraz z procesem przetwarzania sygnałów pochodzących z pomiarów magnetokardiogramów serca płodu.
Rocznik
Strony
147--149
Opis fizyczny
Bibliogr. 7 poz., wykr.
Twórcy
autor
  • Warsaw University of Technology, Faculty of Mechatronics, Institute of Precision and Biomedical Engineering, Poland
autor
  • Warsaw University of Technology, Faculty of Mechatronics, Institute of Precision and Biomedical Engineering, Poland
autor
  • Warsaw University of Technology, Faculty of Mechatronics, Institute of Precision and Biomedical Engineering, Poland
Bibliografia
  • 1. Wakai R. T., Strasburger J. F. , Lizdeal B. J., Gotteiner N. L. (2003) : ,,Magnetocardiographic Rhythm Patterns at lnitiation and Termination of Fetal Supraventricular Tachycardia", Circulation., 21, 107, pp. 307-12.
  • 2. Dunajski Z. , Peters M. J. (1995): ,,Development of the Fetal Magnetocardiograms from the 13PthP Week of Gestation Onward, Proc. of the 9PthP International Conference on Biomagnetism, Vienna, 1995, p. 704-708.
  • 3. Quartero H. W., Stinstra J. G., Golbach E. G., Meijboom E. J ., Peters M. J. (2002): ,,Clinical lmplications of Fetal Magnetocardiography", Ultrasound Obstet Gynecol., 20, pp. 142-53.
  • 4. Vrba J. (2000): ,,Multichannel SQUID biomagnetic systems", In: Weinstock H. (Ed): ,,Applications of superconductivity", (Dordrecht, Kluwer Academic Publishers), pp. 61-138.
  • 5. Malmivuo J., Plonsey R. (1995): ,,Bioelectromagnetism: principles and applications of bioelectric and magnetic fields", (Oxford University Press, New York), pp. 148.
  • 6. Bachir W. , Dunajski Z. , Palko T. (2003): ,,SQUID System with Third Order Gradiometer for Fetal Heart Monitoring", Proc. of WC2003 World Congress on Medical Phys. and Biomed. Eng., Sydney, 2003.
  • 7. Longini R. L., Giolmi J. P., Wall Ili C., and Quick R.F., ,,Filtering without phase shift", IEEE Trans. Biomed. Eng., vol. BME-22, pp. 432-433, 1975.