A 3D localisation system for pre-operating planning is presented. The developed localisation system is able to estimate the 3D position of a target, e.g. an AVM (Arterio-Venous Malformation), based on a set of digital X-ray images. Distortion correction and camera calibrations is performed in order to obtain accurate 3D estimates. The system is tested on a plexiglass phantom with embedded steel pellets offering 126 test points with known 3D positions. The test has shown that the system is able to estimate 3D positions with a mean error of 0.33 mm and a maximum error of 1.18 mm. The required accuracy is 0.16 mm and the test showed that more than 90% of the 3D position estimates fulfilthis requirement.