Application of dose reader CD-07 for depth-dose distribution measurements in the process of microbiological radiation decontamination

Treść / Zawartość
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
Dose reader CD-07 has been introduced for measurements of a dose distribution during dose mapping exercises and for calculations of electron beam energy. This instrument consists of an SP-880 spectrophotometer equipped with a driving motor to move a dosimetric foil across light beam in the front of narrow light aperture. To control the movement of the foil and to present and store the measuring results a specialized software has been developed. In this paper the use of dose reader CD-07 in Performance Qualification of irradiated herbs was described.
Czasopismo
Rocznik
Strony
261--265
Opis fizyczny
Bibliogr. 10 poz., rys.
Twórcy
autor
autor
autor
autor
autor
  • Institute of Nuclear Chemistry and Technology, 16 Dorodna Str., 03-195 Warsaw, Poland, Tel.: +48 22 863 8439, Fax: +48 22 863 9935, sprupr@wp.pl
Bibliografia
  • 1. Directive 1999/2/EC of the European Parliament and of the Council of 22 February on the approximation of the laws of the Member States concerning foods and food ingredients treated with ionizing radiation. Official Journal of the EC (13.3.1999), L66/16-L66/22
  • 2. Chmielewski AG, Sadat T, Zimek Z (2008) Electron accelerators for radiation sterilization. In: Trends in radiation sterilization of health care products. IAEA, Vienna, pp 27–45
  • 3. IAEA (2002) Dosimetry for food irradiation. Technical Report Series no 409. International Atomic Energy Agency, Vienna
  • 4. ISO/ASTM 51649:2005 (2005) Standard practice for dosimetry in an electron beam facility for radiation processing at energies between 300 keV and 25 MeV. International Organization for Standardization, Geneva
  • 5. ISO 11137-3:2006 (2006) Sterilization of health care products – radiation. Part 3. Guidance on dosimetric aspects. International Organization for Standardization, Geneva
  • 6. Jakowiuk A (2007) Operation manual of CD-07 dose reader program. Report INCT Series C, no 20/III/2007. Institute of Nuclear Chemistry and Technology, Warsaw (in Polish)
  • 7. Jakowiuk A, Bartak J, Machaj B (2008) Dose reader CD-07. In: Annual Report 2007. Institute of Nuclear Chemistry and Technology, Warsaw, pp 145–146
  • 8. Katusin-Razem B, Novak B, Razem D (2001) Microbiological decontamination of biological raw materials and corresponding pharmaceutical products by irradiation. Radiat Phys Chem 62:261–275
  • 9. Migdal W, Walis L, Chmielewski AG (1993) The pilot plant for electron beam food irradiation. Radiat Phys Chem 42:567–570
  • 10.Spectrophotometer model SP-880 (2005) Operation manual. Metertech Inc. Version 1.07