Determination of dosimetry parameters of ADVANTAGETM 103Pd brachytherapy seed using MCNP4C computer code

Treść / Zawartość
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
The IsoAid LLC Inc. has been introduced ADVANTAGETM 103Pd brachytherapy seed in 2006. The aim of this work is to determine the dosimetric characteristics of this seed according to AAPM’s recommendation in TG43-U1 using MCNP4C computer code. The dose rate constant has been determined to be 0.694š0.001 cGy.h–1.U–1. The radial dose function has been calculated at distances from 0.25 to 7 cm. Two-dimensional anisotropy function have been calculated at distances from 0.25 to 7 cm and at angles from 0 to 90 degrees centigrade at 10 degrees centigrade increments. The one-dimensional anisotropy function and anisotropy constant have been also calculated. The anisotropy constant in water has been calculated as 0.872š0.001. The results of this investigation are compared with the results of Meigooni et al. obtained by PTRAN code in 2006 and Sowards results obtained by PTRAN code in 2007. The comparison of the dose rate constant and the one-dimensional anisotropy function obtained from the two codes shows good agreement; also the radial dose function at distances lower than 3 cm and the two-dimensional anisotropy function at angles greater than 20 degrees centigrade are in good agreement. But, for the calculated radial dose function at distances beyond 3 cm, we observed differences between our values and Meigooni et al. and Sowards results. Also, differences between the calculated two-dimensional anisotropy function using the two codes for angles smaller than 20 degrees centigrade are considerable. The differences between the results of MCNP4C and PTRAN codes could be related to the different cross-section data libraries used in these two codes.
Czasopismo
Rocznik
Strony
181--187
Opis fizyczny
Bibliogr. 25 poz., rys.
Twórcy
autor
autor
autor
  • Nuclear Engineering Department, Faculty of Engineering, Science and Research Unit, Islamic Azad University, P. O. Box 14515/775, Tehran, Iran, graisali@aeoi.org.ir
Bibliografia
  • 1. Bernard S, Vynckier S (2005) Dosimetric study of a new polymer encapsulated palladium-103 seed. Phys Med Biol 50:1493–1504
  • 2. Bohm TD, Deluka PM Jr, Dewerd LA (2003) Brachytherapy dosimetry of 125I and 103Pd sources using an updated cross-section library for the MCNP Monte Carlo transport code. Med Phys 30;4:701–711
  • 3. Briesmeister JF (ed) (2000) MCNP4C, a general Monte Carlo N Particle Transport Code System, Version 4C. Los Alamos National Laboratory, LA-13709-M
  • 4. Cutrer LM, Hollywood N Calif (1999) Laser welded brachytherapy source and method of making the source.Patent no U S P 5, 997463:1–14
  • 5. Everett CJ, Cashwell ED (1973) MCP code fluorescence--routine revision. Los Alamos National Laboratory Report, LA-5240-MS
  • 6. Hubbell JH, Viegele WJ, Briggs EA, Brown RT, Cromer DT, Howerton RJ (1975) Atomic form factors, incoherent scattering functions, and photons scattering cross sections. J Phys Chem Ref Date 4:471–494
  • 7. Khan FM (1995) The physics of tadiation therapy. Philadelphia, Lippincott William & Wilkins
  • 8. Li Z, Palta JR, Fan JJ (2000) Monte Carlo calculation and experimental measurements of dosimetry parameters of a new 103Pd source. Med Phys 27:1108–1112
  • 9. Mainegra E, Capote R, Lopez E (2000) Radial dose function for 103Pd, 125I, 169Yb, and 192Ir brachytherapy sources: an EGS4 Monte Carlo study. Phys Med Biol 45:703–717
  • 10. Meigooni AS, Bharucha Z, Yoe-Sein M, Sowards K (2001)Dosimetric characteristics of the best double-wall 103Pd brachytherapy source. Med Phys 28:2568–2575
  • 11. Meigooni AS, Dini ShA, Awan ShB, Dou K, Koona RA (2006) Theoretical and experimental determination of dosimetric characterization for ADVANTAGETM Pd-103 brachytherapy source. Appl Radiat Isot 64:881–887
  • 12. Meigooni AS, Sowards K, Soldano M (2000) Dosimetric characteristics of the InterSource103 palladium brachytherapy source. Med Phys 27:1093–1100
  • 13. Meigooni AS, Zhang H, Clark JR, Rachabatthula V, Koona RA (2004) Dosimetric characteristics of the new RadioCoilTM 103Pd wire line source for use in permanent brachytherapy source. Med Phys 31;11:3095–3105
  • 14. Nath R, Anderson LL, Liuxton G, Weaver KA, Williamson JF, Meigooni AS (1995) Report of AAPM radiation therapy committee task group 43. Dosimetry of interstitial brachytherapy sources. Reprinted from Med Phys 22;2:209–234
  • 15. Nath R, Anderson LL, Meli JA, Olch AJ, Stitt JA, Williamson JF (1997) Code of practice for brachytherapy physics: report of the AAPM Radiation Therapy Committee Task Group no 56. Med Phys 24;10:1557–1598
  • 16. Patel NS, Chiu-Tsao S, Williamson JF et al. (2001) Thermo-luminescent dosimetry of the Symmetra 125I model I25.S06 interstitial brachytherapy seed. Med Phys 28;8:1761–1769
  • 17. Reniers B, Vynckier S, Scalliet P (2001) Dosimetric study of the new InterSource 125iodine seed. Med Phys 28:2285–2288
  • 18. Reniers B, Vynckier S, Scalliet P (2002) Dosimetric study of a new palladium seed. Appl Radiat Isot 57:805–811
  • 19. Rivard MJ (2001) Monte Carlo calculation of AAPM Task Group report no 43 dosimetry parameters for the MED3631-A/M 125I source. Med Phys 28;4:629–637
  • 20. Rivard MJ (2001) A discredited approach to determining TG-43 brachytherapy dosimetry parameters: case study using Monte Carlo calculation for the 103Pd MED3633 source. Appl Radiat Isot 55:775–782
  • 21. Rivard MJ, Coursey BM, DeWerd LA et al. (2004) Update of AAPM Task Group no 43 report: a revised AAPM protocol for brachytherapy dose calculations. Med Phys 31;3:633–674
  • 22. Rivard MJ, Melhus ChS, Kirk BL (2004) Brachytherapy dosimetry parameters calculated for a new 103Pd source. Med Phys 31;9:2466–2470
  • 23. Sowards KT (2007) Monte Carlo dosimetric characterization of the IsoAid ADVANTAGE 103Pd brachytherapy source. J Appl Clinical Medical Physics 8;2:18–25
  • 24. Storm E, Israel HI (1967) Photon cross sections from 0.001 to 100 MeV for elements 1 through 100. Los Alamos Scientific Laboratory Report, LA-3753
  • 25. Williamson JF (2000) Monte Carlo modeling of the transverse-axis dose distribution of the model 200 103Pd interstitial brachytherapy seed. Med Phys 27:643–654
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BUJ7-0008-0020
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.