A comparison of the basic photon and electron dosimetry data for Neptun 10PC linear accelerators

Treść / Zawartość
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
. In recent years the similarity of dosimetric characteristics of modern linear accelerators with the same make, model and nominal energy, has become more common. The goal of this study was to quantitatively investigate the reproducibility of the basic photon and electron dosimetry data from Neptun 10PC accelerators across the institutions. In the current study, the photon and electron dosimetry data collected during acceptance and initial commissioning of six Neptun 10PC linear accelerators are analyzed. The dates of original installations of these six machines were evenly spread out over a 5 year period and the series of measurements were conducted during an average of 1-2 months after original installations. All units had identical energies and beam modifiers. For photon beams, the collected data include depth dose data, output factors and beam profile data in water. For electron beams, in addition to depth dose data and output factors, the effective source skin distance for 10 × 10 cm field size is also presented. For most beam parameters the variation (one standard deviation), was less than 1.0% (less than 2% for 2 parameters). A variation of this magnitude is expected to be observed during annual calibration of well-maintained accelerators. In conclusion, this study is presenting a consistent set of data for Neptun 10PC linear accelerators. This consistency implies that for this model, a standard data set of basic photon and electron dosimetry could be established, as a guide for future commissioning, beam modeling and quality assurance purposes.
Czasopismo
Rocznik
Strony
181--185
Opis fizyczny
Bibliogr. 7 poz., rys.
Twórcy
autor
autor
  • Department of Medical Physics and Engineering, School of Medicine, Isfahan University of Medical Sciences, Isfahan, Iran, 81746-73461, Tel.: +98 913 315 8618, Fax: +98 311 668 8597, Shokrani@med.mui.ac.ir
Bibliografia
  • 1. Cho SH, Ibbott GS (2003) Reference photon dosimetry data: a preliminary study of in-air off-axis factor, percentage depth dose, and output factor of the Siemens Primus linear accelerator. J Appl Clin Med Phys 4:300–306
  • 2. Cho SH, Vassiliev ON, Lee S, Liu HH, Ibbott GS, Mohan R (2005) Reference photon dosimetry data and reference phase space data for the 6 MV photon beam from Varian Clinac 2100 series linear accelerators. Med Phys 32:137–148
  • 3. IAEA (2000) Absorbed dose determination in external beam radiotherapy: an international code of practice for dosimetry based on standards of absorbed dose to water. Technical Reports Series no. 398. International Atomic Energy Agency, Vienna
  • 4. Khan FM, Doppke KP, Hogstrom KR (1991) Clinical electron beam dosimetry: report of AAPM Radiation Therapy Committee Task Group No. 25. Med Phys 18:73–109
  • 5. Palta JR, Ayyanar K, Daftari I, Suntheralingam V (1990) Characteristics of photon beams from Philips SL25 linacs. Med Phys 7:106–116
  • 6. SINS (1998) Neptun 10PC. Operating manual. Soltan Institute for Nuclear Studies, Otwock-Świerk, Poland
  • 7. Song H, Xiao Y, Galvin JM (2002) Comparison of characteristics photon and electron beams generated by Philips/Elekta and Varian linear accelerators. Med Phys 29:960–966
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BUJ6-0023-0040
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.