Operation monitoring and automatic control of heart supporting system require information about ventricular assist device output. One of non-invasive measurement methods is pneumatic' chamber resonant frequencies analyse. The acoustic noise brought inside pneumatic chamber is transformed by amplification or suppression of its frequency components. In consequence of this the received signal spectrum contains modified components of stimulation noise. Selected spectrum's regions parameters (like its maximal amplitude value or location along frequency axis) correlate with changes of pneumatic' chamber volume.
Czasopismo
Rocznik
Tom
Strony
89--96
Opis fizyczny
Bibliogr. 13 poz., rys.
Twórcy
autor
autor
autor
autor
- Foundation for Cardiac Surgery Development, 345a Wolnosci Str., Zabrze, POLAND
Bibliografia
- 1.A. Szczeklik. Internal medicine Vol. 1. Cardiovascular system diseases. Practical Medicine Publisher. Warsaw, 2005 (in Polish)
- 2.J. Li. Dynamics of the vascular system. World Scientific Publishing. Singapore, 2004.
- 3.W. A. Missfeldt. Pneumatic drive circuit for an artificial ventricle including systolic pressure control. United States Patent No. 4,808,088. USPT Office, 1989.
- 4.B. Candelon, J. Trinkl. Method and device for flow control of a periodic flow heart prosthesis. European Patent Office No. WO9112035. 1991.
- 5.Novel method to determine instantaneous blood volume in pulsatile blood pomp using electrical impedance. Artificial Organs Vol. 18 No 8, Osaka, 1994.96 Gawlikowski M., Darłak M., Pustelny T., Kustosz R.
- 6.M. Darłak. Master's thesis. Investigations of optical and acoustical methods application for ventricular assist device monitoring. Gliwice, 2003. (in Polish).
- 7.M. Darłak, M. Gawlikowski, R. Kustosz, S. Opilski, T. Pustelny. Patent application nr P-364126. Method and system for temporary blood volume measurement in ventricular assist device, particularly in pneumatically supplied ventricular assist device. Warsaw, 2003. (in Polish).
- 8.M. Gawlikowski, R. Kustosz. Blood volume measurement methods in diaphragm pneumatic heart assist pump. ESAO Annular Meeting. Vienna, 2002.
- 9.Veco Vansonic. Technical documentation. Slim plastic cone speakers. Model No. 11CL32G-M25NT. www.veco.com.tw
- 10.Veco Vansonic. Technical documentation. Electret Condenser Microphones. Model No. PVM-4530-1. www.veco.com.tw
- 11.S. J. Quaal. Cardiac mechanical assistance beyond balloon pumping. Mosby-Year Book, St. Louis, 1997.
- 12.S. Konturek. Physiology of human. Circulatory system. Jagiellonian University Publishing. Cracow, 2001. (in Polish).
- 13.M. Darłak, Z. Opilski, T. Pustelny, M. Gawlikowski, R. Kustosz. Acoustic method of non-invasive blood volume monitoring in POLVAD pump. ESAO Annular Meeting. Bologna, 2005.