Tytuł artykułu

Blood Pressure Measurement Method Using Pulse-Transit-Times

Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
This paper describes a method of measuring blood preasure based on pulse-transit-time. Transducer array based on electro-mechanical film material was employed to detect pulsations non-invasively on the radial artery, while its blood flow was blocked by an occluding cuff. Venous blood preasure was measured simultaneously from a vein on the same hand, and into-arterial blood preasure was obtained from the radial artery on the other hand by means of a catheter tip. Blood preasure was measured in 16 healthy young persons in supine and sitting position during increasing and decreasing cuff preasure. Correlation calculations were performed to assess the performance of the transducer array and the accuracy of the measurement.
Rocznik
Tom
Strony
169--181
Opis fizyczny
Bibliogr. 34 poz., rys.
Twórcy
autor
  • University of Oulu, Department of Electrical and Information Engineering, Optoelectronics and Measurement Techniques Laboratory, P. O. BOX 4500, FIN-90014 University of Oulu, Finland
autor
  • University of Oulu, Department of Electrical and Information Engineering, Optoelectronics and Measurement Techniques Laboratory, P. O. BOX 4500, FIN-90014 University of Oulu, Finland
autor
  • University of Oulu, Department of Electrical and Information Engineering, Optoelectronics and Measurement Techniques Laboratory, P. O. BOX 4500, FIN-90014 University of Oulu, Finland
  • Department of Intemal Medicine and Biocenter Oulu, University of Oulu, 90220 Oulu, Finland
autor
  • Polar Electro Oy, Professorintie 5,90440 Kempele, Finland
autor
  • Polar Electro Oy, Professorintie 5,90440 Kempele, Finland
Bibliografia
  • 1. S. Terry, J. Eckerle, R. Kombluh, T. Low, C. Ablow, “ Silicon pressure transducer arrays for blood pressure measurement,” Sensors and Actuators, A21-A23 (1990), pp. 1070-1079, 1990.
  • 2. L.A. Geddes, Pulse arrival time as a method of obtaining systolic and diastolic blood pressure indirectly. Med. Biol Eng. Comput.,19,671 -672, 1981.
  • 3. W. Chen, T. Kobayashi, S. Ichikawa, Y. Takeuchi, T. Togawa. Continuous estimation of systolic blood pressure using the pulse arrival time and intermittent caiibration. Med.Biol. Eng.Comput, Vol. 38, no. 5, pp. 569-574,2000.
  • 4, Asmar, R., A. Benetos, J. Topouchian, P Laurent., B pannier, A-M. Brisac, R. Target, B. I. Levy. Assessment of arterial distensibility by automatic pulse wave velocity measurement. Hypertension, 26(3): 485-490. 1995.
  • 5. J. Kerola, V. Kontra, R. Sepponen. Non-invasive blood pressure data acquisition employing pulse transit time detection. 18th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Amsterdam 1996, 5.3.1: Vessel Dynamics, pp. 1308-1309, Jan. 1997.
  • 6. Santic, M. Saban. Two methods for determination of diastolic pressures in fingers. 1995 IEEE- EMBC and CMBEC, Theme 1, Cardiovascular System, pp. 149-150, July 1997.
  • 7. Santic, I. Lacrovic. Simultaneous application of multiple oscillometric methods for blood pressure measurement in finger. Proceedings of The First Joint BMES/EMBS Conference Serving Humanity, Advancing Technology, Atlanta, GA, USA, p. 231, Oct. 1999.
  • 8. M. Maguire, T. Ward, C. Markham, D. 0’Shea, L. JCevin. A comparative study in the use of brachial photoplethymography and the QRS complex as timing reference in determination of pulse transit time. In: Engineering in Medicine and Biology Society, 2001. Proceedings of the 23rd Annual International Conference of the IEEE, Vol. 1,2001, pp. 215-218,2001.
  • 9. S. Nissilfi, M. Sorvisto, H. Sorvoja, E. Vieri-Gashi, R. Myllyla. Noninvasive blood pressure measurement based on the electronic palpation method. In: Proc. of the 20th Annual International Conference of the IEEE Engineering in Medicine and Biology Society, Vol. 20, No 4, pp. 1723-1726, 1998.
  • 10. H. Sorvoja, R. Myllylla, S. Nisslla, P. Karja-Koskenkari, J.Koskenkari, M. Lilja, A. Kesfiniemi. A method to determine diastolic blood pressure based on pressure pulse propagation in the electronic palpation method. In: Engineering in Medicine and Biology Society, Proc. of the 23rd Annual International Conference of the IEEE, Vol. 1,2001, pp. 3255-3258,2001.
  • 11. H. Soryoja, R. Myllyla, S. Nissilla, P. Karja-Koskenkari, J. Koskenkari, M. Lilja, A. Kesflniemi. Systolic blood pressure accuracy enhancement in the electronic palpation method using pulse waveform In: Engineering in Medicine and Biology Society, Proc. of the 23rd Annual Internationa] Conference of the IEEE, Vol. I, 2001, pp. 222 -225,2001.
  • 12. E. Vieri-Gashi, H. Sorvoja, R. Myllyla, S. Nissila, M. Sorvisto, P. Karja-Koskenkari. The effect of the venous pressure to the biood pressure signals measured by the palpation method. In: Engineering in Medicine and Biology Society, Proc. of the 23"* Annual International Conference of the IEEE, Vol. ł, 2001, pp. 3259-3261,2001.
  • 13. Vieri-Gashi, E. I., P. Karja-Koskenkari, S. M. Nissilfi, M. Sorvisto, H. S. S. Sorvoja, R. A. Myllyla. Blood pressure variation measured by the electronic paipation method and compared to intra-arterial variation. In: Proceedings of Isik 2000 Workshop on Biomedical Information Engineering, 25-27. June, Istanbul, Turkey, pp. 101-103,2000.
  • 14. K. Meigas, R. Kattai, J. Lass. Continuous blood pressure monitoring using pulse wave delay. Im Engineering in Medicine and Biology Society, Proc. of the 23rd Annual International Conference of the IEEE, Vol. 4, pp. 3171-3174,2001.
  • 15. H. Sarvoja, R. Myllyla. Accuracy of the electronic palpation method. The 74 Conference of the European Society for Engineering and Medicine (ESEM), p. 196,2003.
  • 16. H. Alexander, M. Cohen, L. Steinfeld. Criteria in the choice of an occluding cuff for the indirect measurement of the blood pressure. Med.Bio. Eng.Comput., 15:2,1977.
  • 17. L. A, Geddes, S. J. Whistler. The error in indirect blood pressure measurement with incorrect size of cuff. Am Heart J, 96:4, 1978.
  • 18. L.A. Geddes, L. A. Handbook of Blood Pressure Measurement Humana Press. 1991. pp. 51-61.
  • 19. Yarows, S. A., S. Julius, T.G. Pickering, Home Blood Pressure Monitoring. Arch Intern Med/Vol. 160, May 8, pp. 1251-1257,2000,
  • 20. Association for the Advancement of Medicai Instrumentation. American national standard. Electronic or automated sphygmomanometers. ANS1/AAM1 SP 10-1987, Arlington, VA: AAMI, 1987:25.
  • 21. Association for the Advancement of Medicai Instrumentation. American national standard. Electronic or automated sphygmomanometers. ANS1/AAMI SP 10-1992, Arlington, VA: AAMI, 1993:1-40.
  • 22. E. 0’Brien, J. Petrie, W. Littler, et al. The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring device with special reference to ambulatoiy systems. J.Hyperterts, Vol. 8, pp, 607-619,1990.
  • 23. E. O’Brien, J. Petrie, W. Littler, et al. Short report: an outłine of the revised British Hypertension Society protocol for the evaluation blood pressure measuring devices. J Hypertens, Vot. 11, pp. 677-679,1993.
  • 24. O’Brien, E., B, Waeber, G. Parati, J Staessen, M. G. Myers. Clinical Review: Blood pressure measuring devices: recommendations of the European Society of Hypertension. BMJ, Vol. 322, pp. 531-536,2001.
  • 25. K. Kirjavainen. Electomechanicat film and procedure for manufacturing same. US Patent 4,654,546,9 p., March 1987.
  • 26. L. RfiisSnen, R. Pohjavirta, M. Unkila, J. Tuomisto, A new method for the measurement of spontaneous motor activity of laboratory animals. Phormacology & Toxicology, vol. 70, pp. 230-231. 1992.
  • 27. J. Siivola, K. Leinonen, L. Rilis&nen, ETMF-polymer transducer as a detector of respiration in humans. Med & Biot Etjg & Comput., vol. 31, pp. 634-635,1993.
  • 28. J. Lekkala, R. Poramo, K. Nyholm, T. Kaikkonen. EMF force sensor - a flexible and sensitive electret film for physiological applications. Med & Biol Eng& Comput., vol. 34, Suppl. 1, Part 1, pp, 67-68, 1996.
  • 29. M. Paajanen, H. Valimaki, J. Lekkala. Modelling the eiectromechanical film (EMFi). J Ełectrostatics, vol. 48 (2000), pp. 193-204.2000.
  • 30. J. Peitonen, M. Paajanen, J. Lekkala. Determination of the actuator sensitivity of eiectromechanical polypropylene films by atomic force. Am. J. Appl. Phys,, vol. 88, No. 8, pp. 4769-4793,2000.
  • 31. M. Paajanen, J. Lekkala, K. Kirjavainen. Eiectromechanical film (EMFi)- a new multipurpose electret materiał. Sensors and Actuators, vol. 84 (2000), pp. 95-102.
  • 32. M. Paajanen, M. Wegener, R, Gerhard-Multhaupt. Understanding the role o!f the gas in the voids during corona charging of cellular electret films- a way to enhance their pietzoetectricity. J. Phys. D: Appl Phys., vol. 34(2001), pp. 2482-2488,2001.
  • 33. M. Paajanen, J. Lekkala, H. Wlimfiki. Eiectromechanical modeling and properties of the electret film EMFI, IEEE Transactions on Dielectrics and Electrical Insulation, vol. 8 No. 4, pp. 629-636,2001.
  • 34. S. Nissila, Miettinen Jari. Measuring Arrangement. European Patent Application EP 1034736 Al, 2000.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BUJ6-0008-0040
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.