A gauge was developed for determining the dose rate distribution and surface activity of ophthalmic brachytherapy applicators, particularly for 106Ru applicators. A plastic fi 2×2 mm scintillator is used as the radiation detector, featuring a high pulse count rate, which results in a low 0.5% random error, due to good counting statistics. Automatic gain control of the photomultiplier tube (PMT) is achieved using a LED as the reference light source. The PMT operates in pulse mode. Long term gain variation due to fatigue of the PMT or ambient temperature variation is thus compensated for. The count rate error due to inaccurate setting of the high voltage supply of the PMT is 0.4%, and the instability error over 7 hours of continuous operation does not exceed 1-2%, peak-to-peak.
Słowa kluczowe
Czasopismo
Rocznik
Tom
Strony
107--111
Opis fizyczny
Bibliogr. 3 poz., rys.
Twórcy
autor
- Institute of Nuclear Chemistry and Technology, 16 Dorodna Str., 03-195 Warsaw, Poland, Tel.: +48 22/ 811 06 55, Fax: +48 22/ 811 19 17
autor
- Institute of Nuclear Chemistry and Technology, 16 Dorodna Str., 03-195 Warsaw, Poland, Tel.: +48 22/ 811 06 55, Fax: +48 22/ 811 19 17
autor
- Institute of Nuclear Chemistry and Technology, 16 Dorodna Str., 03-195 Warsaw, Poland, Tel.: +48 22/ 811 06 55, Fax: +48 22/ 811 19 17
autor
- Health Physics Laboratory, Institute of Nuclear Physics, 152 Radzikowskiego Str., 31-342 Kraków, Poland
autor
- Health Physics Laboratory, Institute of Nuclear Physics, 152 Radzikowskiego Str., 31-342 Kraków, Poland
Bibliografia
- 1.De Almeida C, DeWerd L, Jarvinen H, Soares CG (2000) Guidelines for the calibration of low energy photon sources and beta-ray brachytherapy sources. IAEA Report No. DMRP-200002.IAEA, Vienna
- 2.Evans RD (1970) The atomic nucleus. McGraw-Hill Book Company, New York, Toronto, London
- 3.Fluhs D, Heintz M, Wieczorek C (1996) Direct reading measurement of absorbed dose with plastic scintillators. The general concept and applications to ophthalmic plaque. Med Phys 23:427–435
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BUJ6-0006-0042