X-ray tube with needle-like anode

Treść / Zawartość
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
An X-ray tube with a needle-like anode (NAXT) built in our Laboratory, its design and basic operating parameters are presented. The process of electron beam forming and influence of external and internal magnetic fields is discussed. The tube properties essential from the point of view of its application in X-ray generators as well as disadvantageous thermal effects caused by flow of heat generated in the tube target to irradiated objects are discussed. The tube is almost a point-like source of X radiation emitted into 4đ geometry; the dose rates are on the order of 1 Gy/min at the distance of 10 mm from the anode cap. Preliminary tests show the tube may be useful in brachytherapy of cancer tumors of diameter up to 30 mm. The tube may also be an interesting device in widely understood field of irradiation techniques.
Słowa kluczowe
Czasopismo
Rocznik
Strony
101--105
Opis fizyczny
Bibliogr. 5 poz., rys.
Twórcy
autor
  • The Andrzej Sołtan Institute for Nuclear Studies, 05-400 Otwock-Swierk, Poland, Tel.: +48 22/ 718 05 61, Fax: +48 22/ 779 34 81
autor
  • The Andrzej Sołtan Institute for Nuclear Studies, 05-400 Otwock-Swierk, Poland, Tel.: +48 22/ 718 05 61, Fax: +48 22/ 779 34 81
autor
  • The Andrzej Sołtan Institute for Nuclear Studies, 05-400 Otwock-Swierk, Poland, Tel.: +48 22/ 718 05 61, Fax: +48 22/ 779 34 81
autor
  • Power Dora System, 110 S. Wyszynskiego Str., 50-307 Wroclaw, Poland
autor
  • The Andrzej Sołtan Institute for Nuclear Studies, 05-400 Otwock-Swierk, Poland, Tel.: +48 22/ 718 05 61, Fax: +48 22/ 779 34 81
autor
  • Uniquant 14/9 Batalionu Parasol Str., 01-118 Warsaw, Poland
autor
  • The Andrzej Sołtan Institute for Nuclear Studies, 05-400 Otwock-Swierk, Poland, Tel.: +48 22/ 718 05 61, Fax: +48 22/ 779 34 81
Bibliografia
  • 1. Dinsmore M, Harte KJ, Sliski AP et al. (1996) A new miniature X-ray source for interstitial radiosurgery: device description. Med. Phys 23;1:45–52
  • 2. Dyson NA (1973) X-rays in atomic and nuclear physics. Longman Group Limited, London
  • 3. Hakim R, Zervas NT, Hakim F et al. (1997) Initial characterization of the dosimetry and radiobiology of a device for administering interstitial stereotactic radiosurgery. Neurosurgery 40;3:510–516
  • 4. Skillicorn B (1982) X-ray tubes for energy dispersive XRF spectrometry. Advance in X-Ray Analysis 25:49–57
  • 5. Yanch C, Zervas NT (1995) The photon radiosurgery system. Sci Am Sci Med 2;6:38–47
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BUJ6-0006-0041
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.