PL EN


Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników
Tytuł artykułu

Identification of mechanical parameters of respiratory system during ventilatory support of lungs

Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
Identification of mechanical parameters of lungs by using a pressure method with an added compliance is presented in this paper. This method is based on periodically connecting an additional external mechanical compliance to the respirator-lungs circuit, to determine the values of the mechanical parameters of the respiratory system from measurements of the pressure in specific phases of the respiratory cycle. The expiratory flow was measured to estimate the time constant and the respiratory resistance. The identification of the parameters of lungs is connected with the dynamic signal analysis of the pressure courses registered in a very short time (< 200 ms) at the beginning of the expiratory phase, when the respiratory muscles are non-active. The measurement system is driven by the computer, which is also used to analyse the signals received from the lung model. The results confirm that the measurement time of 200 ms is sufficient to determine the lung compliance.
Twórcy
autor
  • Institute of Biocybernetics and Biomedical Engineering, Polish Academy of Sciences, 02-109 Warszawa, ul. Ks. Trojdena 4
autor
  • Institute of Biocybernetics and Biomedical Engineering, Polish Academy of Sciences, 02-109 Warszawa, ul. Ks. Trojdena 4
autor
  • Institute of Biocybernetics and Biomedical Engineering, Polish Academy of Sciences, 02-109 Warszawa, ul. Ks. Trojdena 4
Bibliografia
  • [1] Doboszyńska A., Wrotek K.: Badania czynnościowe układu oddechowego, PZWL, Warszawa 1999.
  • [2] Nunn J.F.: Kliniczna fizjologia oddychania, PZWL, Warszawa 1981.
  • [3] Darowski M., Kozarski M., Gólczewski T.: Model Studies on Respiratory Parameters for Different Lung Structures, Biocybemetics and Biomedical Engineering, 20, 2, 2000, 67-77.
  • [4] Darowski M., Symulacja procesu wspomagania oddychania i sztucznej wentylacji płuc, IBIB, PAN, Warszawa 1997.
  • [5] Darowski M., Gottlieb-Inacio I., Ludwigs U. and Hedenstiema G.: Assessment of respiratory system compliance by a flow recording method. Acta Anaesthesiologica Scand, 1995, 39, 462-466.
  • [6] Darowski M., Jucha A., Kozarski M., Pałko K.J.: Analiza krzywej wydechowej dla oceny mechaniki płuc metodą dołączonej podatności, XII Krajowa Konferencja Naukowa Biocybernetyka i Inżynieria Biomedyczna, Warszawa 2001, tom 1, 252-258.
  • [7] Darowski M., Pałko K. J., Jucha A., Kozarski M.: Expiratory pressure curve analysis for estimation of lungs mechanics, Biocybernetics and Biomedical Engineering, 23, 1, 2003, 91-102.
  • [8] Kozarski M., Darowski M., Gumiężny R.: Metoda ciśnieniowa dołączonej podatności w ocenie parametrow mechanicznych płuc, XI Krajowa Konferencja Naukowa Biocybernetyka i Inżynieria Biomedyczna, Warszawa 1999, tom 1, 197-202.
  • [9] Felicio Da Silva Jose: Monitorage des properties mechaniques thoracopulmonaires au cours de la ventilation artificielle, PhD thesis, De L’lnstinstitut Polytechnique de Lorraine (I.N.P.L), Nancy 1992.
  • [10] M. Kirshner, S. Katz: Design theory of fluid components, New York, Academic Press, 1975, ISBN 0124102506.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPZ3-0006-0068
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.