Based on Laplace's law and the mechanical characteristics of knitted vascular prostheses, mathematical models have been developed enabling one to determine the values of the relative circumferential elongation, circumferential forces and standard circumferential susceptibility of a prosthesis as a function of its diameter and internal pressure. The results obtained refer to the values of the parameters assessed for the initial phase of their exploitation, allow to preliminarily predict the values of these parameters.
PL
Na podstawie prawa Laplace'a i charakterystyki mechanicznej dzianiny stosowanej do wytwarzania protez naczyń krwionośnych opracowano modele matematyczne umożliwiające wyznaczenie wartości obwodowego wydłużenia względnego, sił obwodowych i normatywnej podatności obwodowej w funkcji średnicy protezy i ciśnienia wewnętrznego. Wyniki uzyskane z tych analiz odnoszą się do wartości ocenianych parametrów dla początkowej fazy ich eksploatacji i pozwolą one na wstępne prognozowanie wartości tych parametrów.
Rocznik
Strony
87--91
Opis fizyczny
Bibliogr. 14 poz.
Twórcy
autor
autor
autor
  • Poland, Łódź, Technical University of Lodz, Department of Knitting Technology
Bibliografia
  • 1. Norma ISO 7198:1998 (Cardiovascular implants – Tubular vascular prostheses).
  • 2. Goossens D, et al. In vivo size of knitted Dacron prostheses (Gelseal) used in the thoracic aorta: a computed tomography study, Cardiovasc. Surg. 1996; 4(2): 205-206.
  • 3. Mattens E. Gelseal versus Gelweave dacron prosthestic grafts in the descending thoarcic aorta: a two-year computed tomography scan follow-up study. Cardiovasc. Surg. 1999; 7(4): 432–435.
  • 4. Goëau-Brissonnière OA, Qanadli SD, Ippoliti A, Pistolese GR, Coggia M, Pollock JG. Can knitting structure affect dilation of polyester bifurcated prostheses? A randomized study with the use of helical computed tomography scanning. J Vasc Surg. 2000; Jan 31(1 Pt 1): 157.
  • 5. Vermeulen F, et al. The Hemashield Woven prosthesis in the thoracic aorta: a prospective computer tomography follow-up study, Cardiovascular Surgery 2001; 9: 6: 580–585.
  • 6. Bogaert J, Dymarkowski S, Budts W, Gewillig M, Daenen W. Graft dilation after redo surgery for aneurysm formation following patch angioplasty for aortic coarctation, Eur. J. Cardiothorac Surg. 2001; 19: 274 -278.
  • 7. Schepens MAAM. Dilation of knitted grafts in the thoracic aorta, European Journal of Cardio-thoracic Surgery 2001; 20:430.
  • 8. Utoh J, Goto H, Hirata T, Hara M, Kitamura N. Dilation of sealed Dacron vascular prostheses: a comparison of Gelseal and Hemashield. J Cardiovasc Surg 1998; 39: 179-180.
  • 9. Riepe G, Chafke N, Morlock M, Imig H. Dilatation and durability of polyester grafts. In: Branchereau A, Jacobs M (eds) Complications in vascular and endovascular surgery, part I. Armonk, NY: Futura Publishing, 2001.
  • 10. Nabil Chakfe, et al. In Vitro Approach to the Dilative Behavior of Knitted Vascular Prosthetic Grafts, Ann Vasc Surg 2008; 22: 402-411.
  • 11. Ocena podatności obwodowej protez naczyń krwionośnych na dynamiczne zmiany ciśnienia wewnętrznego podczas badań zmęczeniowych in vitro. Projekt badawczy nr 0411/R/P01/2008/04.
  • 12. Jasińska I, Kowalski K, Tarnowski W, Mielicka E, Struszczyk M. Assessment of Circumferential Dynamic Compilance of Knitted Vascular Prothesis in vitro Using a Long Last Fatique. In: FiberMed11, Tampere, Finland, 28-30 June 2011.
  • 13. Korliński W. Podstawy Dziewiarstwa, WNT, Warszawa, 1976.
  • 14. Mikołajczyk Z. Analysis of Warp Knitted Fabrics Relative Cover with the use of Computer Technique of Image Processing, Fibres & Textiles in Eastern Europe 2001, 9; 3 (34): 26-29.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPW7-0023-0005
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.