Electrical substitution cryogenic radiometer based spectral responsivity scale between 250–2500 nm wavelengths

Autorzy
Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
Języki publikacji
EN
Abstrakty
EN
This paper presents the spectral responsivity scale between 250–2500 nm developed by the National Metrology Institute of Turkey (UME). For that purpose silicon photodiode based trap detector and electrically calibrated pyroelectric radiometer (ECPR) that were calibrated against primary level absolute electrical substitution cryogenic radiometer (ESCR) were used as transfer standards. Using highly collimated and stabilized (10–5) lasers, absolute optical powers and absolute responsivity of trap detector were measured with an uncertainty of the order of 10–4. In visible (VIS) region responsivity scale was set by means of the models for the reflectance and internal quantum efficiency. In the ultraviolet (UV) and near-infrared (NIR) regions spectrally flat (0.1%) ECPR was used.
Czasopismo
Rocznik
Strony
427--438
Opis fizyczny
Bibliogr. 27 poz., rys.
Twórcy
autor
  • TUBITAK-Ulusal Metroloji Enstitüsü (UME), P.O. Box. 54 41470 Kocaeli/Turkey
autor
  • TUBITAK-Ulusal Metroloji Enstitüsü (UME), P.O. Box. 54 41470 Kocaeli/Turkey
Bibliografia
  • [1] MARTIN J.E., FOX N.P., KEY P.J., Metrologia 21 (1985), 147.
  • [2] OHNO Y., CROMER C.L., HARDIS C.E., EPPELDAUER G., 23 (1994), 88.
  • [3] GOODMAN T.M., KEY P.J., Metrologia 25 (1988), 29.
  • [4] BOIVIN L.P., GAERTNER A.A., GIGNAC D.S., Metrologia 24 (1987), 139.
  • [5] TOIVANEN P., KARHA P., MANOOCHEHRI F.,IKONEN E., Metrologia 37 (2000), 131.
  • [6] TIMO VARPULA, HEIKKI SEPPA, JUHA-MATTI SARRI, IEEE Transactions On Instrumentation and Measurement 38 (1989), 558.
  • [7] ZALEWSKI E.F., DUDA C.R., Appl. Opt. 19 (1983), 2867.
  • [8] FOX N.P., Metrologia 28 (1991), 197.
  • [9] GEIST J., Phys. Rev. B 27 (1983), 4841.
  • [10] KORDE R., GEIST J., Appl. Opt. 26 (1987), 5284.
  • [11] GARDNER J.L., BROWN W.J., Appl. Opt. 26 (1987), 2431.
  • [12] DOYL W.M., MCINTOSH B.C., GEIST J., Opt. Eng. 15 (1976), 541.
  • [13] KÖHLER R., GOEBEL R., PELLOR., Metrologia 33 (1996), 549.
  • [14] GENTILE T.R., HOUSTON J.M., HARDIS J.E., CROMER C.L., PARR A.C., Appl. Opt. 35 (1996), 1056.
  • [15] STOCK K.D., HOFER H., Metrologia 30 (1993), 291.
  • [16] NETTLETON D.H., PRIOR T.R., WARD T.H., Metrologia 30 (1993), 425.
  • [17] GENTILE T.R., HOUSTON S.M., CROMER C.L., Appl. Opt. 35 (1996), 4392.
  • [18] DURANT N.M., FOX N.P., Metrologia 30 (1993), 345.
  • [19] BOIVIN L.P., MCNEELY F.T., Appl. Opt. 25 (1986), 554.
  • [20] HAAPALINNA A., KARHA P., IKONEN E., Appl. Opt. 31 (1998), 729.
  • [21] KARHA P., FAGERLUND H., LASSILA A., LUDVIGSEN H., MONOOCHEHRI F., IKONEN E., Opt. Eng. 34 (1995), 2611.
  • [22] VARPULA T., SEPPA H., SAARI J.-M., IEEE Transactions on Instrumentation and Measurement 38 (1989), 558.
  • [23] PALMER J.M., Metrologia 30 (1993), 327.
  • [24] ZALEWSKI E.F., HOYT C.C., Metrologia 28 (1991), 203.
  • [25] GEIST J., Appl. Opt. 18 (1979), 760.
  • [26] WERNER L., FISCHER J., JOHANNSEN U., HARTMANN J., Metrologia 37 (2000), 279.
  • [27] LAASON T.C., BRUCE S., PARR A.C., NIST Special Publication 250-41, 1998, p. 29