Tytuł artykułu
Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
Planarne czujniki z tlenku rutenu w "cell-on-a-chip" do badań metabolicznych
Języki publikacji
Abstrakty
While the use of indium oxide electrodes for neuronal stimulation and pH sensing is continuously increasing, ruthenium oxide is a material less familiar in biomedicine. However, ruthenium oxide deposited by reactive sputtering glass or ceramic substrate with a platinum contact layer is shown to yield sensors well suitable for a extracellular recording of cell-mediated pH changes in cell culture media. Ruthenium oxide was directly grown with MCF-7 and L 929 cells and is fully biocompatible. The pH sensitivity ranges between 52 and 58 mV/pH. it is approximately linear between pH 5.5 and pH 11. The drift typically ranges between I and 2 mV h-1 With increasing thickness of the ruthenium oxide layer, drift decreases and response time increases. A practicable tradeoff is a thickness of 1-2 um. A pronounced redox cross-sensitivity of ruthenium oxide is a limiting factor in situations where concentration of dissolved oxygen is not constant or cannot be determined independently. The observed sensitivity to dissolved oxygen is ≈ 0.2 mV/hPa. Ruthenium oxide spots on ceramic sensor chips were used for an exemplary cell based assay with MCF-7 cells and showed a response of extracellular acidification rate to the alkaloid drug cytocha-lasin B.
Zastosowanie elektrod wykorzystujących tlenek irydu w neuronowej stymulacji i pomiarach pH wzrasta, lecz tlenek rutenu jest mniej powszechnie spotykany w biomedycynie. Niemniej, wykazano, że tlenek rutenu osadzony w wyniku napylania na szklanym lub ceramicznym substracie z kontaktem platynowym może tworzyć czujniki przydatne do pozakomórkowego rejestrowania regulowanych przez komórkę zmian pH w kulturach komórkowych. Tlenek rutenu jest w pełni kompatybilny z komórkami MCF-7 i L 929. Czułość pH waha się w granicach miedzy 52 a 58 mV/pH, i jest w przybliżeniu liniowa między pH 5,511. Dryft zwykle jest zawarty w granicach l do 2 mV h-1. Wzrost grubości warstwy tlenku rutenu powoduje zmniejszenie dryftu i przedłużenie czasu odpowiedzi. Optymalna grubość wartswy wynosi 1-2 urn. Interferencje redoks są istotnym czynnikiem ograniczającym stosowanie czujników gdy stężenie rozpuszczonego tlenu jest zmienne lub gdy nie może być niezależnie określone. Zaobserwowana czułość na rozpuszczony tlen wynosi ≈ 0,2 mV/hPa. Kropki tlenku rutenu w ceramicznych czujnikach były stosowane z komórkami MCF-7 i pozwaliły obserwować wpływ leku cytochalazyny B na szybkość pozakomórkowego zakwaszania.
Wydawca
Czasopismo
Rocznik
Tom
Strony
1193--1201
Opis fizyczny
Bibliogr. 23 poz.
Twórcy
autor
autor
autor
autor
autor
autor
- Heinz Nixdorf-Lehrstuhl fur Medizinische Elektronik TU Munchen, D-80333 Munchen, Germany Fax: +49 89 289 22950, brischwein@tum.de
Bibliografia
- 1. Cooper M.A., Drug Discovery Today, 11, 1068 (2006).
- 2. Ges I.A., Ivanov B.L., Werdich A.A. and Baudenbacher F.J., Biosens. Bioelectron., 22, 1303 (2007).
- 3. Walker G.M., Zeringue H.C. and Beebe D.J., Lab Chip, 4, 91 (2004).
- 4. Wiest J., Stadthagen T., Schmidhuber M., Brischwein M., Ressler J., Raeder U., Grothe H., Melzer A. and Wolf B., Anal. Lett., 39, 1759 (2006).
- 5. Gatenby R.A. and Gillies R.J., Int. J. Biochem. Cell Biol., 39, 1358 (2007).
- 6. Harguindey S., Orive G., Pedraz J.L., Paradiso A., and Reshkin S.J., Biochim. Biophys. Acta, 1756, 1 (2005).
- 7. Henning T., Kraus M., Brischwein M., Otto A.M. and Wolf B., Anti-Cancer Drugs, 15, 7 (2004).
- 8. McConnell H.M., Owicki J.C., Parce J.C., Miller D.L., Baxter G.T., Wada H.G. and Pitchford S., Science, 257, 1906 (1992).
- 9. Poghossian A., Ingebrandt S., Offenhäuser A. and Schöning M.J., Semin. Cell Dev. Biol., 20, 41 (2009).
- 10. Ceriotti L., Kob A., Drechsler S., Ponti J., Thedinga E., Colpo P., Ehret R. and Rossi F., Anal. Biochem., 371, 92 (2007).
- 11. Ressler J., Sensorchips für die multiparametrische zelluläre Bioanalytik und für biohybride Bauelemente, PhD Thesis at Technische Universität München (2008).
- 12. O.Hare D., Parker K.H. and Winlove C.P., Med. Eng. Phys., 28, 982 (2006).
- 13. Glab S., Hulanicki A., Edwall G. and Ingman F., Crit. Rev. Anal. Chem., 21, 29 (1989).
- 14. Koncki R. and Mascini M., Anal. Chim. Acta, 351, 143 (1997).
- 15. Fog A. and Buck R., Sens. Actuators, 5, 137 (1984).
- 16. Brischwein M., Motrescu E.R., Otto A.M., Cabala E., Grothe H. and Wolf B., Lab Chip, 3, 234 (2003).
- 17. Olthuis W., Robben M.A.M., Bergveld P., Bos M. and Van Der Linden W.E., Sens. Actuators B, 2, 247 (1990).
- 18. Healy T.W., Yates D.E., White L.R. and Chan D., J. Electroanal. Chem., 80, 57 (1977).
- 19. Wang, M., Yao S. and Madou M., Sens. Actuators B, 81, 313 (2002).
- 20. Hendrikse J., Olthuis W. and Bergveld P., Sens. Actuators B, 53, 97 (1998).
- 21. Ferrick D.A., Neilson A. and Beeson C., Drug Discovery Today, 13, 268 (2008).
- 22. Geisler T., Ressler J., Harz H., Wolf B. and Uhl R., IEEE Trans. Autom. Sci. Eng., 3, 169 (2006).
- 23. Yang H., Kang, S.K., Choi, C.A., Kim H., Shin D.-H., Shin Kim Y. and Tae Kim, Y., Lab Chip, 4, 42 (2004).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPP2-0003-0025