PL EN


Preferencje help
Widoczny [Schowaj] Abstrakt
Liczba wyników
Tytuł artykułu

Automated sequential monitoring of ammonium, phosphate and nitrite in wastewater by multi-commutated peristaltic and solenoid pumped flow system : a comparative study

Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
PL
Automatyczny, sekwencyjny monitoring jonów amonowych, fosforanowych i azotynowych w ściekach za pomocą wielodrożnego perystaltycznego oraz solenoidowego pompowego układu przepływowego : badanie porównawcze
Języki publikacji
EN
Abstrakty
EN
A comparative study on analytical performance of a basic multi-commutated solenoid pump flow system (MPFS) and multi-commutated FIA system (MCFS) has been performed. Both flow systems utilised three-port solenoid valves acting as fluidic switches. The MCFS system was obtained by replacing a solenoid pump with a peristaltic pump, synchronized with valves, as a propelling unit. These two continuous flow strategies were applied to spectro-photometric determination of ammonium, nitrite, and phosphate in wastewater samples from urban wastewater treatment plant. Optimised peristaltic and solenoid pumped flow systems yielded similar results with respect to sensitivity', linear range . MPFS was. however, superior to MCFS in terms of day-to-day reproducibility (RSDs calculated for the slopes of 20 calibration curves obtained in different days were: , sample consumption (reduced twice for determination of ammonium and nitrite), and sample throughput . MPFS was advantageous with respect to versatility, possibility of miniaturisation, and increased simplicity due to reconfiguration of single-commutated elements without the need of physical reconfiguration of the flow manifold.
PL
Wykonano badania porównawcze sprawności analitycznej wielodrożnego przepływowego układu pompy solenoidowej (MPFS) i wielodrożnego układu wstrzykowej analizy przepływowej (MCFS). W obu układach zastosowano trójdrożne zawory solenoidowe działające jako przełączniki strumieniowe. Układ MCFS otrzymano przez zastąpienie pompy solenoidowej pompą perystaityczną, zsynchronizowaną/ zaworami, jako jednostką napędową. Te dwie strategie ciągłego przepływu zastosowano do spektrofotometrycznego oznaczania lonów amonowych, azotynów i fosforanów w próbkach ścieków z oczyszczalni miejskiej. Zoptymalizowane układy pompowe przepływu - perystaltyczny i solenoidowy dały podobne wyniki w odniesieniu do czułości, zakresu liniowości . MPFS był jednak lepszy niż MCFS w przypadku odtwarzalności z dnia na dzień (względne odchylenie standardowe obliczone dla 20 wykresów kalibracyjnych wyznaczonych w różnych dniach wynosiło: zużycia próbki (mniejszego dwukrotnie przy oznaczaniu NH4: i NO2 oraz liczby analiz (86 h-1 MPFS był również lepszy ze względu na wszechstronność, możliwość miniaturyzacji i większą prostotę - możliwość zamiany jcdnodrożnych elementów bez konieczności fizycznej rekonstrukcji przewodów.
Czasopismo
Rocznik
Strony
757--770
Opis fizyczny
Bibliogr. 29 poz.
Twórcy
  • Departamento de Quimica, Bioquimicay Biologia Molecular, Universidad Cardenal Herrera-C.E.U., 46113 Moncada (Valencia), SPAIN Fax: 0034 96 139 52 72, jvgarcia@uch.ceu.es
Bibliografia
  • 1.Ruzicka J. and Marshall G.D., Anal. Chim. Acta, 237, 329 (1990).
  • 2.Hansen E.H., Talanta, 64, 1076 (2004).
  • 3.Reis B.F., Giné M.F., Zagatto E.A.G., Lima J.L.F.C. and Lapa R.A.S., Anal. Chim. Acta, 293,129 (1994).
  • 4.Rothwell S.D. and Woolf A. A., Talanta, 32,431 (1985).
  • 5.Albertus F., Horstkotte B., Cladera A. and Cerdá V., Analyst, 124, 1373 (1999).
  • 6.Rocha F.R.P., Martelli P.B. and Reis B.F.. Anal. Chim. Acta, 438. 11 (2001).
  • 7.Dasgupta P.K., Vithanage R.S. and Petersen K., Anal. Chim. Acta, 215,277 (1988).
  • 8.Weeks D.A. and Johnson K.S., Anal. Chem.. 68. 2717 (1996).
  • 9.Rocha F.R.P., Torralba E.R., Rubio A.M. ad de la Guardia M., Anal. Chim. Acta, 547,204 (2005).
  • 10.Pons C., Forteza R. and Cerdá V., Anal. Chim. Acta, 550, 33 (2005).
  • 11.Lapa R.A.S., Lima J.L.F.C., Reis B.F., Santos J.L.M. and Zagatto E.A.G., Anal. Chim. Acta, 466, 125 (2002).
  • 12.Francis P.S., Lewis S.W.. Lim K.F., Calisson K. and Karlberg B., Ta/anta, 58, 1029 (2002).
  • 13.Lima J.L.F.C., Santos J.L.M.. Dias A.C.B.. Ribeiro M.F.T. and Zagatto E.A.G., Talanta, 64, 1091 (2004).
  • 14.Santos J.L.M., Ribeiro M.F.T., Dias A.C.B., Lima J.L.F.C., Zagatto E.A.G., Santos J.L. and Lima J.L.F.C., Anal. Chim. Acta, in press. Available on line 21 January (2007). Presented as Plenary Lecture at the X International Conference of Flow Analysis, 3-8 September (2006) Porto, Portugal.
  • 15.Hoke S.H., Anal. Chim. Acta, 460,219 (2002).
  • 16.Carneiro J.M.T., Zagatto E.A.G., Santos J.L.M. and Lima J.L.F.C., Anal. Chim. Acta, 474, 161 - (2002).
  • 17.LimaJ.L.F.C., Sá S.M.O., Santos J.L.M. and Zagatto E.A.G.,/. Pharm. Biomed. Anal., 32, 1011 (2003).
  • 18.Dias A.C.B, Santos J.L.M., LimaJ.L.F.C. and Zagatto E.A.G., Anal. Chim. Acta, 499,107 (2003).
  • 19.Carneiro J. M.T., Dias A.C.B., Zagatto E. A. G., Santos J.L.M. and Lima J.L.F.C .,Anal. Chim. Acta,531, 279 (2005).
  • 20.Marques K.L., Santos J.L.M. and Lima J.L.F.C., Anal. Bioanal. Chem., 382,452 (2005).
  • 21.Chapin T.P. and Wanty R.B., Anal. Chim. Acta, 543. 199 (2005).
  • 22.Lewis S.W., Francis P.S., Lim K.F., Jenkins G.E. and Wang X.D., The Analyst, 125, 1869 (2000).
  • 23.Pires C.K., Marques K.L., Santos J.L.M., Lapa R.A.S., LimaJ.L.F.C. and Zagatto E.A.G., Talanta,68, 239 (2005).
  • 24.Lavorante A.F., Rubio A.M., de la Guardia M. and Reis B.F., Anal. Bioanal. Chem., 381, 1305(2005).
  • 25.”Métodos Normalizados para el Análisis de Aguas Potables y Residuales”, [Díaz de Santos S. A.,Ed.], American Public Health Association, American Water Works, Association Water Polution Control Federation (APHA-AWW-AWP). 20th edn, Madrid 1992.
  • 26.Santos M.M., Dosreis B.F., Bergamin F.H. and Baccan N., Anal. Chim. Acta, 261, 339 (1992) .
  • 27.Legnerova Z., Solich P., Sklenarova H., Satinsky D. and Karlicek R„ Water. Res.,36,2777(2002).
  • 28.Gabriel D., Baeza J., Valero F. and Lafuente J., Anal. Chim. Acta, 359, 173 (1998).
  • 29.Auflitsch S., Peat D.M.W., Mckelvie I. and Worslfold P.J., Analyst, 122, 1477 (1997).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPP1-0079-0005
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.