Fast gas chromatography of polar and non-polar compounds with the on-column injection

Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
PL
Szybka chromatografia gazowa polarnych i niepolarnych związków z bezpośrednim dozowaniem do kolumn
Języki publikacji
EN
Abstrakty
EN
Fast gas chromatography (GC) with the on-column injection of large volumes of polar compounds of a broad range of volatility (in the range of volatilities ofrt-alkanes C16-C26) has been performed. The GC system comprised a normal-bore retention gap (1 m x 0.32 mm I.D. and/or 0.53 mm I.D.) coupled with a narrow-bore analytical column (5 m x 0.1 mm I.D. x 0.4 [am film thickness, HP-1 MS) using a standard glass press-fit connector. The effect of the injection volume, pre-column I.D. and polarity, and solvent polarity and volatility on the peak width and shape, as well as on the focusing of polar pesticides and n-alkanes has been discussed. The precision of chromatographic measurements with hydrogen as the carrier gas under the constant inlet pressure and fast temperature programming for fast separations of large volume injections (1-8 μL) was found to be good. For polar pesticides the occurrence of adsorption phenomena was evidenced. They were dependent on the quality of the pre-column.
PL
Za pomocą szybkiej chromatografii gazowej, stosując bezpośrednie dozowanie do kolumny dużych objętości, chromatografowano związki polarne różniące się znacznie lotnością (w zakresie lotności n-alkenów C16-C26). W chromatografie zastosowano przedkolumnę, I m x 0,32 mm i/lub 0.53 mm, połączoną z kolumną analityczną 5 m x O, l mm x 0.4 μm; HP-1 MS. Przedyskutowano wpływ objętości zadozowanej próbki, średnicy i polarności przed kolumny oraz lotności i polarności rozpuszczalnika na szerokość i kształt piku oraz na szerokość pasm polarnych pestycydów i n-alkenów. Stwierdzono, że precyzja pomiarów chromatograficznych przy zastosowaniu wodoru jako gazu nośnego, stałego ciśnienia na wlocie oraz szybkiego programowania temperatury do szybkiego rozdzielania dużych objętości (1-8 1-8 μL) była dobra. W przypadku polarnych pestycydów wystąpiły zjawiska adsorpcyjne. Były one zależne od jakości przedkolumny.
Czasopismo
Rocznik
Strony
365--375
Opis fizyczny
Bibliogr. 19 poz.
Twórcy
  • Slovak University of Technology. Faculty of Chemical and Food Technology, Department of Analytical Chemistry, Radlinského 9, 812 37 Bratislava, Slovak Republic
autor
  • Slovak University of Technology. Faculty of Chemical and Food Technology, Department of Analytical Chemistry, Radlinského 9, 812 37 Bratislava, Slovak Republic
autor
  • Slovak University of Technology. Faculty of Chemical and Food Technology, Department of Analytical Chemistry, Radlinského 9, 812 37 Bratislava, Slovak Republic
autor
  • Slovak University of Technology. Faculty of Chemical and Food Technology, Department of Analytical Chemistry, Radlinského 9, 812 37 Bratislava, Slovak Republic
Bibliografia
  • 1. Korytár P., Janssen H.-G., Matisová E. and Brinkman U.A.Th., Trends Anal. Chem., 21, 558 (2002).
  • 2. Matisová E. and Dömötörová M., J. Chromatogr. A, 1000, 199 (2003).
  • 3. Cramers C.A. and Leclercq P.A., J. Chromatogr. A, 842, 3 (1999).
  • 4. Guichon G., Anal. Chem., 50, 1812 (1978).
  • 5. Korytár P., Matisová E., Lefflerová H. and Slobodník J., J. High. Resol. Chromatogr., 23, 149 (2000).
  • 6. Matisová E., Šimeková M., Hrouzková S., Korytár P. and Dömötörová M., J. Sep. Sci., 25, 1325 (2002).
  • 7. Dagan S. and Amirav A., J. Am. Soc. Mass Spectrom., 7, 737 (1996).
  • 8. Cochran J.W., J. Chromatogr. Sci., 40, 254 (2002).
  • 9. Dallűge J., Ou-Aissa R., Vreuls J.J. and Brinkman U.A.Th., J. High Resol. Chromatogr., 22, 459 (1999).
  • 10. Maštovská K., Lehotay S.J. and Hajšlová J., J. Chromatogr. A, 926, 291 (2001).
  • 11. Maštovská K., Hajšlová J., Godula M., Křivánková J. and Kocourek V., J. Chromatogr. A, 907, 235 (2001).
  • 12. Jackson G.P. and Andrews A.R.J., Analyst, 123, 1085 (1998).
  • 13. Jager L.S. and Andrews A.R.J., Analyst, 125, 1943 (2000).
  • 14. Grob K. Jr., J. High Resol. Chromatogr. Commun., 1, 461 (1984).
  • 15. Grob K., On-Column Injection in Capillary Gas Chromatography, 2nd edn, Hütig, Heidelberg 1991, p. 362.
  • 16. Dömötörová M., in: Engineer Diploma Thesis, Slovak University of Technology, Faculty of Chemical and Food Technology, Department of Analytical Chemistry, Bratislava, 2002.
  • 17. Grob K., On-column Injection in Capillary Gas Chromatography, 2nd edn, Hütig, Heidelberg 1991, p. 125.
  • 18. Matisová E., Kakalíková E., Leško J. and de Zeeuw J., J. Chromatogr. A, 754, 445 (1996).
  • 19. J. Ďurčanská (personal communications).1. Korytár P., Janssen H.-G., Matisová E. and Brinkman U.A.Th., Trends Anal. Chem., 21, 558 (2002).
  • 2. Matisová E. and Dömötörová M., J. Chromatogr. A, 1000, 199 (2003).
  • 3. Cramers C.A. and Leclercq P.A., J. Chromatogr. A, 842, 3 (1999).
  • 4. Guichon G., Anal. Chem., 50, 1812 (1978).
  • 5. Korytár P., Matisová E., Lefflerová H. and Slobodník J., J. High. Resol. Chromatogr., 23, 149 (2000).
  • 6. Matisová E., Šimeková M., Hrouzková S., Korytár P. and Dömötörová M., J. Sep. Sci., 25, 1325 (2002).
  • 7. Dagan S. and Amirav A., J. Am. Soc. Mass Spectrom., 7, 737 (1996).
  • 8. Cochran J.W., J. Chromatogr. Sci., 40, 254 (2002).
  • 9. Dallűge J., Ou-Aissa R., Vreuls J.J. and Brinkman U.A.Th., J. High Resol. Chromatogr., 22, 459 (1999).
  • 10. Maštovská K., Lehotay S.J. and Hajšlová J., J. Chromatogr. A, 926, 291 (2001).
  • 11. Maštovská K., Hajšlová J., Godula M., Křivánková J. and Kocourek V., J. Chromatogr. A, 907, 235 (2001).
  • 12. Jackson G.P. and Andrews A.R.J., Analyst, 123, 1085 (1998).
  • 13. Jager L.S. and Andrews A.R.J., Analyst, 125, 1943 (2000).
  • 14. Grob K. Jr., J. High Resol. Chromatogr. Commun., 1, 461 (1984).
  • 15. Grob K., On-Column Injection in Capillary Gas Chromatography, 2nd edn, Hütig, Heidelberg 1991, p. 362.
  • 16. Dömötörová M., in: Engineer Diploma Thesis, Slovak University of Technology, Faculty of Chemical and Food Technology, Department of Analytical Chemistry, Bratislava, 2002.
  • 17. Grob K., On-column Injection in Capillary Gas Chromatography, 2nd edn, Hütig, Heidelberg 1991, p. 125.
  • 18. Matisová E., Kakalíková E., Leško J. and de Zeeuw J., J. Chromatogr. A, 754, 445 (1996).
  • 19. J. Ďurčanská (personal communications).
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPP1-0049-0056
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.