A simultaneous spectrometer with photodiode array detector and low power radiofrequency capacitively coupled plasma source

Wybrane pełne teksty z tego czasopisma
Identyfikatory
Warianty tytułu
PL
Spektrometr do jednoczesnego oznaczania z mozaikowym detektorem i źródłem wzbudzenia pojemnościowo sprzężonej plazmy małej mocy
Języki publikacji
EN
Abstrakty
EN
The main figures of merit of a simultaneous spectrometer with photodiode array (PDA) detector and low power (275 W; 27.12 MHz) and Ar consumption (0.4 L min(-1)) radiofre-quency capacitively coupled plasma source for simultaneous determination of some alkaline and alkaline-earth elements (Li, Na, K, Ca) are presented. The torch has a molybdenum tubular electrode and a single (SRT) or double ring electrode (DRT) distanced at 60 mm. The simultaneous spectrometer equipped with PDA detector (200-1100 nm) has a 95 mm Paschen-Runge mount. For the best geometry of the plasma (DRT configuration), the limits of detection are: 30 (Li); 40 (Na); 60 (K) ng mL(-1) and 210; 100; 210; 90 ng mL(-1) for Ca(II)393.3 nm, Ca(I) 422.7 nm, CaOH 554 nm and CaOH 622 nm. The recovery degree of Ca in certified plasma sera was 99.4š1.4 % and RSD 3.8-4,9%. In the determination of Ca in five blood samples the RSD was 2.9-4.4%.
PL
Zaprezentowano wielkości charakterystyczne spektrometru z detektorem mozaikowym (PDA) i źródłem wzbudzenia plazmy malej mocy (275 W; 27.12 MHz) przy zużyciu argonu 0.4 L min(-1), do jednoczesnego oznaczania niektórych pierwiastków alkalicznych i ziem alkalicznych (Li, Na, K, Ca). Palnik wyposażony był w molibdenową elektrodę rurową i jedno- (SRT) lub dwupierścieniową (DRT) elektrodę umieszczoną w odległości 60 mm. Równoczesny spektrometr z detektorem PDA (200-1100 nm) pracuje w 95 nm układzie Paschena-Runge'a. Przy najlepszej geometrii plazmy (konfiguracja DRT) granice detekcji wynosiły: 30 (Li); 40 (Na); 60 (K) ngmL(-1) oraz 210; 100:210; 90 ugraL(-1) dla Ca(II) 393.3 nm, Ca(I) 422.7 nm, CaOH 554 nm i CaOH 622 nm. Odzysk Ca w certyfikowanych surowicach wyniósł 99.4š1.4 % a względne odchylenie standardowe 3.8-4.9%. Podczas oznaczania Ca w pięciu próbkach krwi względne odchylenie standardowe wyniosło 2.9-4.4%.
Czasopismo
Rocznik
Strony
725--736
Opis fizyczny
Bibliogr. 17 poz.
Twórcy
autor
  • Babes-Bolyai University Cluj-Napoca, Faculty of Chemistry, Arany Janos 11, 3400 Cluj-Napoca, Romania
autor
  • Babes-Bolyai University Cluj-Napoca, Faculty of Chemistry, Arany Janos 11, 3400 Cluj-Napoca, Romania
autor
  • Babes-Bolyai University Cluj-Napoca, Faculty of Physics, M. Kogalniceanu 1, 3400 Cluj-Napoca, Romania
autor
  • Babes-Bolyai University Cluj-Napoca, Faculty of Physics, M. Kogalniceanu 1, 3400 Cluj-Napoca, Romania
autor
  • Babes-Bolyai University Cluj-Napoca, Faculty of Chemistry, Arany Janos 11, 3400 Cluj-Napoca, Romania
autor
  • Babes-Bolyai University Cluj-Napoca, Faculty of Chemistry, Arany Janos 11, 3400 Cluj-Napoca, Romania
Bibliografia
  • 1. Blades M.W., Banks P., Gill C., Huang D. and Liang D.C., IEEE Trans. Plasma Sci., 19, 1090 (1991).
  • 2. Blades M.W., Spectrochim. Acta, Part B, 49B, 47 (1994).
  • 3. Sturgeon R.E., Willie S.N., Luong V.T. and Berman S.S., J. Anal. At. Spectrom., 4, 669 (1989).
  • 4. Liang D.C. and Blades M.W., Spectrochim. Acta, Part B, 44B, 1059 (1989).
  • 5. Gilchrist G.F.R., Cellier P.M., Yang H., Yu C. and Liang D.C., J. Anal. At. Spectrom., 8, 809 (1993).
  • 6. Cordos E.A., Anghel S.D., Frentiu T. and Popescu A., J. Anal. At. Spectrom., 9, 635 (1994).
  • 7. Frentiu T., Rusu A.M, Ponta M., Anghel S.D. and Cordos E.A., Fresenius J. Anal. Chem., 355, 254 (1996).
  • 8. Frentiu T., Anghel S.D., Simon A., Popescu A., Rusu A.M. and Cordos E.A., ACH Models in Chem., 136, 131 (1999).
  • 9. Anghel S.D., Frentiu T., Cordos E.A., Simon A. and Popescu, A., J. Anal. At. Spectrom., 14, 541 (1999).
  • 10. Cordos E.A., Frentiu T., Rusu A.M., Anghel S.D., Fodor A. and Ponta, M., Talanta, 48, 827 (1999).
  • 11. Frentiu T., Ponta M., Anghel S.D., Simon A. and Cordos E.A., Anal. Letters, 33, 323 (2000).
  • 12. Frentiu T., Anghel S.D., Nicola M., Darvasi E., Simon A. and Cordos E.A., Croat. Chem. Acta, 72, 763 (1999).
  • 13. Frentiu T., Rusu A.M., Anghel S.D., Negoescu S., Popescu A., Simon A. and Cordos E.A., ACH Models in Chem., 136, 119 (1999).
  • 14. Skoog D.A., Holler F.J. and Nieman T.A., Principles of Instrumental Analysis, 5th edn, Saunders College Publishing, 1998, p. 200.
  • 15. Dean J.A., in: Flame photometry, McGrow-Hill Book Company, Inc., New York, 1960, p. 153-203.
  • 16. James C.G. and Sugden T.M., Nature, 175, 333 (1955).
  • 17. Miller J.C. and Miller J.N., in: Statistics for Analytical Chemistry, 2 nd edn, John Wiley and Sons, New York 1988; pp. 53-62, 120-124.
Typ dokumentu
Bibliografia
Identyfikator YADDA
bwmeta1.element.baztech-article-BPP1-0031-0007
JavaScript jest wyłączony w Twojej przeglądarce internetowej. Włącz go, a następnie odśwież stronę, aby móc w pełni z niej korzystać.